Transgastric, pulsed Doppler echocardiographic determination of cardiac output
- PMID: 8857435
- DOI: 10.1007/BF01728328
Transgastric, pulsed Doppler echocardiographic determination of cardiac output
Abstract
Objective: The aim of this study was to evaluate the accuracy of cardiac output measurement with transesophageal echocardiography (TEE) using a transgastric, pulsed Doppler method in acutely ill patients.
Design: Cardiac output was simultaneously measured by thermodilution (TD) and a transgastric, pulsed Doppler method.
Setting: The study was carried out in a surgical intensive care unit as part of the management protocol of the patients.
Patients: Thirty consecutive acutely ill patients with a Swan-Ganz catheter, mechanically ventilated, sedated and with a stable hemodynamic condition were included.
Measurements: Pulsed Doppler TEE was performed using a transgastric approach in order to obtain a long axis view of the left ventricle. Cardiac output was calculated from the left ventricular outflow tract diameter, the velocity time integral of the blood flow profile and heart rate.
Results: One patient was excluded because of the presence of aortic regurgitation and another, because of the impossibility of obtaining a transgastric view. Twenty-eight simultaneous measurements were performed in 28 patients. A clinically acceptable correlation and agreement were found between the two methods (Doppler cardiac output = 0.889 thermodilution cardiac output +0.74 l/min, r = 0.975, p <0.0001).
Conclusion: Transgastric pulsed Doppler measurement across the left ventricular outflow tract with TEE is a very feasible and clinically acceptable method for cardiac output measurement in acutely ill patients.
Similar articles
-
Pulsed Doppler transesophageal echocardiographic determination of cardiac output in human beings: comparison with thermodilution technique.Am Heart J. 1993 Oct;126(4):956-62. doi: 10.1016/0002-8703(93)90712-i. Am Heart J. 1993. PMID: 8213455
-
Noninvasive cardiac output monitoring by aortic blood flow determination: evaluation of the Sometec Dynemo-3000 system.Crit Care Med. 1998 Dec;26(12):2066-72. doi: 10.1097/00003246-199812000-00043. Crit Care Med. 1998. PMID: 9875922
-
Comparing Methods for Cardiac Output: Intraoperatively Doppler-Derived Cardiac Output Measured With 3-Dimensional Echocardiography Is Not Interchangeable With Cardiac Output by Pulmonary Catheter Thermodilution.Anesth Analg. 2018 Aug;127(2):399-407. doi: 10.1213/ANE.0000000000002800. Anesth Analg. 2018. PMID: 29324489
-
[Noninvasive measurement of cardiac output by pulsed Doppler echocardiography. Correlation with thermodilution].Arch Mal Coeur Vaiss. 1990 Feb;83(2):237-44. Arch Mal Coeur Vaiss. 1990. PMID: 2106860 Review. French.
-
Hemodynamic assessment of critically ill patients using echocardiography Doppler.Curr Opin Crit Care. 2005 Jun;11(3):227-34. doi: 10.1097/01.ccx.0000159946.89658.51. Curr Opin Crit Care. 2005. PMID: 15928471 Review.
Cited by
-
Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.PLoS One. 2019 Oct 3;14(10):e0222105. doi: 10.1371/journal.pone.0222105. eCollection 2019. PLoS One. 2019. PMID: 31581196 Free PMC article.
-
Assess the clinical applicability of transthoracic echocardiography in laparotomy patients in a general intensive care unit.J Res Med Sci. 2014 May;19(5):398-403. J Res Med Sci. 2014. PMID: 25097620 Free PMC article.
-
Clinical monitoring of cardiac output assessed by transoesophageal echocardiography in anaesthetised dogs: a comparison with the thermodilution technique.BMC Vet Res. 2017 Nov 9;13(1):325. doi: 10.1186/s12917-017-1227-9. BMC Vet Res. 2017. PMID: 29121915 Free PMC article.
-
Systematic review of cardiac output measurements by echocardiography vs. thermodilution: the techniques are not interchangeable.Intensive Care Med. 2016 Aug;42(8):1223-33. doi: 10.1007/s00134-016-4258-y. Epub 2016 Mar 1. Intensive Care Med. 2016. PMID: 26932349