Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients
- PMID: 8874306
- DOI: 10.1097/00003246-199610000-00014
Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients
Abstract
Objective: To evaluate the accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients.
Design: Prospective data collection.
Setting: Emergency department in a 2,000-bed hospital.
Patients: Thirty-nine patients admitted to the emergency department requiring invasive arterial blood pressure monitoring were enrolled to the study protocol.
Interventions: Continuous noninvasive blood pressure measurement was performed on the middle phalanx of the second and third finger, using a test instrument which provides continuous arterial waveform display with the use of a finger cuff. Invasive mean arterial blood pressure measurement was done by cannulation of the radial artery and direct transduction of the systemic arterial pressure waveform.
Measurements and main results: Three thousand one hundred eighteen pairs of simultaneous finger cuff and intra-arterial blood pressure measurements were collected in 1-min intervals from 39 patients over a total of 51.8 hrs. The overall discrepancy between both measurements was 0.10 mm Hg. The standard deviation of the differences was +/- 5.02 mm Hg. The mean bias in patients treated with catecholamines was 0.01 mm Hg and was not different from the bias observed in patients without catecholamines (mean bias: 0.23 mm Hg; p > .22). Whereas 95% of all comparisons between finger cuff and intra-arterial measurement had a discrepancy < or = +/- 10 mm Hg, 4.7% had a discrepancy between +/- 10.1 to 15 mm Hg and 0.3% exhibited a discrepancy > +/- 15 mm Hg. In 29 (74%) patients, the duration of errors > 10 mm Hg was < or = 1 min. In seven (18%) patients, the duration of errors > 10 mm Hg was between 2 to 3 mins and in three (8%) patients, the errors lasted for > 3 mins.
Conclusions: Our data provide a guide to the accuracy and reliability of noninvasive finger blood pressure measurements in critically ill patients. Although most test instrument measurements were reliable, in 8% of all patients large discrepancies (> 10 mm Hg) between both measurements with a duration of > 3 mins were noted. Concerning the considerable risk for arterial cannulation, our preliminary data demonstrate that the test instrument (PORTAPRES, TNO Biomedical Instrumentation Research Unit; The Netherlands) is an advance in noninvasive monitoring of critically ill patients and may be useful in most emergency clinical settings.
Similar articles
-
Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients.Crit Care Med. 2000 Feb;28(2):371-6. doi: 10.1097/00003246-200002000-00014. Crit Care Med. 2000. PMID: 10708169
-
Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients.Crit Care Med. 2003 Mar;31(3):793-9. doi: 10.1097/01.CCM.0000053650.12025.1A. Crit Care Med. 2003. PMID: 12626986
-
The reliability of continuous noninvasive finger blood pressure measurement in critically ill children.Anesth Analg. 2009 Mar;108(3):814-21. doi: 10.1213/ane.0b013e318194f401. Anesth Analg. 2009. PMID: 19224788
-
Continuous noninvasive pulse wave analysis using finger cuff technologies for arterial blood pressure and cardiac output monitoring in perioperative and intensive care medicine: a systematic review and meta-analysis.Br J Anaesth. 2020 Jul;125(1):25-37. doi: 10.1016/j.bja.2020.03.013. Epub 2020 May 29. Br J Anaesth. 2020. PMID: 32475686
-
[Noninvasive, continuous measurement of finger artery pressure with the servo-plethysmo-manometer Finapres].Herz. 1989 Aug;14(4):251-9. Herz. 1989. PMID: 2676814 Review. German.
Cited by
-
Increasing sleep duration to lower beat-to-beat blood pressure: a pilot study.J Sleep Res. 2013 Jun;22(3):295-304. doi: 10.1111/jsr.12011. Epub 2012 Nov 22. J Sleep Res. 2013. PMID: 23171375 Free PMC article. Clinical Trial.
-
Positive airway pressure improves nocturnal beat-to-beat blood pressure surges in obesity hypoventilation syndrome with obstructive sleep apnea.Am J Physiol Regul Integr Comp Physiol. 2016 Apr 1;310(7):R602-11. doi: 10.1152/ajpregu.00516.2015. Epub 2016 Jan 27. Am J Physiol Regul Integr Comp Physiol. 2016. PMID: 26818059 Free PMC article. Clinical Trial.
-
The mean machine; accurate non-invasive blood pressure measurement in the critically ill patient.J Clin Monit Comput. 2009 Oct;23(5):283-97. doi: 10.1007/s10877-009-9195-7. Epub 2009 Aug 12. J Clin Monit Comput. 2009. PMID: 19672679
-
The mean prehospital machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient.J Clin Monit Comput. 2010 Jun;24(3):191-202. doi: 10.1007/s10877-010-9236-2. Epub 2010 Jun 8. J Clin Monit Comput. 2010. PMID: 20532593
-
Inter-day reliability of blood pressure and cerebral blood flow velocities in persons with spinal cord injury and intact controls.J Spinal Cord Med. 2017 Mar;40(2):159-169. doi: 10.1080/10790268.2015.1135556. Epub 2016 Feb 17. J Spinal Cord Med. 2017. PMID: 26860937 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources