Evaluation of four monoclonal antibodies against HLA-B27 for their reliability in HLA-B27 typing with flow cytometry (FC): comparison with the classic microlymphocytotoxic test (MLCT)
- PMID: 8889393
- DOI: 10.1002/(SICI)1097-0320(19960915)26:3<209::AID-CYTO5>3.0.CO;2-9
Evaluation of four monoclonal antibodies against HLA-B27 for their reliability in HLA-B27 typing with flow cytometry (FC): comparison with the classic microlymphocytotoxic test (MLCT)
Abstract
Typing for HLA-B27 by serological methods is routinely performed using the microlymphocytotoxic test (MLCT). Since monoclonal antibodies (MAB) against HLA-B27 are available, flow cytometry (FC), which requires less time than the MLCT has been developed as an alternative technique. The aim of the present study was to check the accuracy and reliability of this method using different MAB against HLA-B27 in comparison to MLCT (using polyclonal antibodies against HLA-B27 and cross-reacting specificities [CRS]). FC was performed in 144 patients with HLA-B27-related rheumatic disorders (seronegative spondarthritides) using a special software package which requires corresponding calibration beads in order to achieve a standardized setup of the flow cytometer. MAB from the following producers were used: Becton & Dickinson (BD), Behringwerke (BE), One Lambda (OL), and Immunotech (IT). In addition to the critical limit of fluorescence intensity (FI) which indicates positivity if exceeded, provided by the software (but valid only for the MAB from BD), empirically twice the value of the STD calculated from the mean of the FI values of HLA-B27 positive patients was regarded a good cut off for the HLA-B27 positivity in FC measurements with the MAB used. Using a standard protocol including an incubation of whole EDTA-anticoagulated blood for 15 min with MAB against HLA-B27 (FITC-conjugated) and CD3 (PE-conjugated) and a lysis of erythrocytes, good discrimination between HLA-B27 positive and negative patients was obtained. Cross reactions with HLA-B27 positive patients occurred except when the MAB from OL was used. One false-negative result was found with OL's MAB (out of 22) and false-positive results occurred in HLA-B7+ patients when MAB from BD, BE, and IT were used. Unfortunately also 1 false-positive result (out of 57) was obtained in HLA-B7-, B27- patients with IT's MAB. Errors in the interpretation of the FC analysis might be avoided if more than one MAB (including those not cross reacting with HLA-B7) are used.
Similar articles
-
HLA-B27 determination using serological methods. A comparison of enzyme immunoassay and a microlymphocytotoxic test with flow cytometry and a molecular biological assay.Rheumatol Int. 1996;16(3):95-100. doi: 10.1007/BF01409980. Rheumatol Int. 1996. PMID: 8893373
-
[Use of flow cytometry for HLA B27 phenotyping: study of a HLA B7/HLA B27 double marker technique].Allerg Immunol (Paris). 1997 Jan;29(1):11-4. Allerg Immunol (Paris). 1997. PMID: 9213414 French.
-
Evaluation of a method for monoclonal antibody HLA-B27 analysis with the CELL-DYN Sapphire haematology analyser.Int J Lab Hematol. 2007 Dec;29(6):454-60. doi: 10.1111/j.1751-553X.2007.00912.x. Int J Lab Hematol. 2007. PMID: 17988301
-
[HLA-B27 in Bechterew's disease].Tidsskr Nor Laegeforen. 2000 Apr 30;120(11):1317-22. Tidsskr Nor Laegeforen. 2000. PMID: 10868094 Review. Norwegian.
-
Monoclonal antibodies to HLA-B27.Br J Rheumatol. 1983 Nov;22(4 Suppl 2):110-8. doi: 10.1093/rheumatology/xxii.suppl_2.110. Br J Rheumatol. 1983. PMID: 6360285 Review.
Cited by
-
Variability in the interpretation of microsatellite patterns with different electrophoretic conditions.Mol Pathol. 1999 Oct;52(5):302-4. doi: 10.1136/mp.52.5.302. Mol Pathol. 1999. PMID: 10748882 Free PMC article.
-
Usage of Conventional PCR Technology for the Detection of HLA-B27 Allele: A Significant Molecular Marker of Ankylosing Spondylitis.Indian J Clin Biochem. 2013 Apr;28(2):189-92. doi: 10.1007/s12291-012-0261-4. Epub 2012 Sep 26. Indian J Clin Biochem. 2013. PMID: 24426208 Free PMC article.
-
Human HLA-B27 typing using the BD™ HLA-B27 kit on the BD FACSVia™ system: A multicenter study.Cytometry B Clin Cytom. 2018 Sep;94(5):651-657. doi: 10.1002/cyto.b.21630. Epub 2018 Apr 11. Cytometry B Clin Cytom. 2018. PMID: 29476701 Free PMC article.
-
A Review on the Role of Ethylenediaminetetraacetic Acid (EDTA) in the Treatment and Understanding of Psoriasis.Cureus. 2021 Jul 16;13(7):e16424. doi: 10.7759/cureus.16424. eCollection 2021 Jul. Cureus. 2021. PMID: 34277313 Free PMC article. Review.
-
HLA-B*27 typing by sequence specific amplification without DNA extraction.Mol Pathol. 1999 Oct;52(5):300-1. doi: 10.1136/mp.52.5.300. Mol Pathol. 1999. PMID: 10748881 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials