The value of colour Doppler imaging in assessing flow through ventriculo-peritoneal shunts
- PMID: 8891363
- DOI: 10.1007/BF00261623
The value of colour Doppler imaging in assessing flow through ventriculo-peritoneal shunts
Abstract
So far, Doppler ultrasonography has had little application in paediatric neurosurgery. The incidental observation of normally invisible CSF flow from the peritoneal end of a ventriculo-peritoneal (V-P) shunt on colour Doppler during a conventional abdominal ultrasound examination in one of our patients suggested that colour Doppler imaging might be of potential interest. Using a 10-5 MHz broad-band linear array transducer, conventional 2D and Doppler (spectral and colour-coded) ultrasonography was performed in 17 patients with V-P shunts, who were aged 3 months to 12 years. In all, 20 examinations were performed, because 2 patients had repeat examinations before and after shunt revision. In 13 examinations (65%) CSF flow was identified through the shunt tube. Flow velocities between 5 and 7 cm/s were measured. No flow was seen in 7 examinations (35%), 3 of which (15%) were performed before revisions for a blocked shunt. Further in vitro studies with simulated shunt/ventricular system model demonstrated that although clear CSF is not ultrasonically visible, the presence of particulate matter, such as choroid plexus debris, can generate a satisfactory Doppler signal with the probe insonating over the length of the tubing. In the presence of clear CSF, turbulence generated at junction points of the shunt system or at the exit of the peritoneal tube can be visualised well with Doppler ultrasonography, presumably due to generation of microbubbles. The clinical implications for the management of shunt obstruction are discussed.
Similar articles
-
Radiographic markers of clinical outcomes after endoscopic third ventriculostomy with choroid plexus cauterization: cerebrospinal fluid turbulence and choroid plexus visualization.J Neurosurg Pediatr. 2016 Sep;18(3):287-95. doi: 10.3171/2016.3.PEDS15364. Epub 2016 May 13. J Neurosurg Pediatr. 2016. PMID: 27177081 Free PMC article.
-
Noninvasive cerebrospinal fluid shunt flow measurement by Doppler ultrasound using ultrasonically excited bubbles: a feasibility study.Ultrasound Med Biol. 1999 Mar;25(3):371-89. doi: 10.1016/s0301-5629(98)00164-1. Ultrasound Med Biol. 1999. PMID: 10374981 Clinical Trial.
-
Proof of the patent subcutaneous fibrous tract in children with V-P shunt malfunction.Childs Nerv Syst. 2000 Jun;16(6):351-6. doi: 10.1007/s003810050530. Childs Nerv Syst. 2000. PMID: 10933230
-
Abdominal CSF pseudocysts in patients with ventriculo-peritoneal shunts. Report of fourteen cases and review of the literature.Acta Neurochir (Wien). 1994;127(1-2):73-8. doi: 10.1007/BF01808551. Acta Neurochir (Wien). 1994. PMID: 7942187 Review.
-
Pleural effusion from intrathoracic migration of a ventriculo-peritoneal shunt catheter: pediatric case report and review of the literature.Ital J Pediatr. 2018 Mar 27;44(1):42. doi: 10.1186/s13052-018-0480-2. Ital J Pediatr. 2018. PMID: 29587815 Free PMC article. Review.
Cited by
-
Quantitative contrast-enhanced ultrasound measurement of cerebrospinal fluid flow for the diagnosis of ventricular shunt malfunction.J Neurosurg. 2015 Dec;123(6):1420-6. doi: 10.3171/2014.12.JNS141014. Epub 2015 Jun 19. J Neurosurg. 2015. PMID: 26090831 Free PMC article.
-
Sonographic localization of a nonpalpable shunt: Ultrasound-assisted ventricular shunt tap.Surg Neurol Int. 2013 Aug 6;4:101. doi: 10.4103/2152-7806.116151. eCollection 2013. Surg Neurol Int. 2013. PMID: 24032076 Free PMC article.
-
Superb microvascular ultrasound is a promising non-invasive diagnostic tool to assess a ventriculoperitoneal shunt system function: a feasibility study.Neurosurg Rev. 2024 Sep 2;47(1):521. doi: 10.1007/s10143-024-02665-w. Neurosurg Rev. 2024. PMID: 39222164 Free PMC article.
-
Noninvasive CSF shunt patency evaluation by superb microvascular imaging.Neurosurg Rev. 2023 Aug 1;46(1):190. doi: 10.1007/s10143-023-02090-5. Neurosurg Rev. 2023. PMID: 37526749 Free PMC article.
-
Fluid flow measurement for diagnosis of ventricular shunt malfunction using nonlinear responses of microbubbles in the contrast-enhanced ultrasound imaging.Jpn J Appl Phys (2008). 2017 Jul;56(7 Suppl 1):07JF10. doi: 10.7567/JJAP.56.07JF10. Epub 2017 Jun 13. Jpn J Appl Phys (2008). 2017. PMID: 29937574 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical