High-performance liquid chromatographic assay for 25-hydroxyvitamin D3 in serum
- PMID: 890914
High-performance liquid chromatographic assay for 25-hydroxyvitamin D3 in serum
Abstract
We describe a precise and specific method for measuring 25-hydroxyvitamin D3 in 1 ml of human serum. Extraction with chloroform/methanol followed by chromatography on a 0.5 X 2-cm silica gel column yields a sample that is sufficiently free of extraneous material for high-performance liquid chromatography on a column of microporous silica gel (10 micron average particle diameter). The measurement is not influenced by vitamins D2 or D3, 25-hydroxyvitamin D2, or any of the more hydroxylated metabolites of the vitamin D group. Results by this method correlate well with a competitive protein-binding assay (r = 0.961), but with a negative bias of 6.9 +/- 3.3 microgram/liter. We measured concentrations of 25-hydroxyvitamin D3 in serum drawn during February from 24 persons who were judged normal by physical examinations. The range was 5.5-23.8 microgram/liter (mean, 15.0 +/- 5.2 microgram/liter). The day-to-day CV for the assay was 5.46%.
Similar articles
-
Some problems in development of a high-performance liquid chromatographic assay to measure 25-hydroxyvitamin D2 and 25-hydroxyvitamin D3 simultaneously in human serum.Clin Chem. 1978 Jun;24(6):927-30. Clin Chem. 1978. PMID: 657481
-
Assay of vitamins D2 and D3, and 25-hydroxyvitamins D2 and D3 in human plasma by high-performance liquid chromatography.Clin Chem. 1978 Feb;24(2):287-98. Clin Chem. 1978. PMID: 203413
-
Radioligand receptor assay for 25-hydroxyvitamin D2/D3 and 1 alpha, 25-dihydroxyvitamin D2/D3.J Clin Invest. 1976 Jul;58(1):61-70. doi: 10.1172/JCI108459. J Clin Invest. 1976. PMID: 1084355 Free PMC article.
-
Measurement of 25-hydroxyvitamin D3 in serum.Clin Chem. 1976 Mar;22(3):364-8. Clin Chem. 1976. PMID: 3295
-
A sensitive competitive protein binding assay for vitamin D in plasma.Steroids. 1981 May;37(5):581-91. doi: 10.1016/s0039-128x(81)90403-7. Steroids. 1981. PMID: 6266082
Cited by
-
Association of vitamin D and vitamin D receptor gene polymorphisms with chronic inflammation, insulin resistance and metabolic syndrome components in type 2 diabetic Egyptian patients.Meta Gene. 2014 Aug 7;2:540-56. doi: 10.1016/j.mgene.2014.07.002. eCollection 2014 Dec. Meta Gene. 2014. PMID: 25606437 Free PMC article.
-
Influence of Vitamin D Binding Protein on Accuracy of 25-Hydroxyvitamin D Measurement Using the ADVIA Centaur Vitamin D Total Assay.Int J Endocrinol. 2014;2014:691679. doi: 10.1155/2014/691679. Epub 2014 Jun 19. Int J Endocrinol. 2014. PMID: 25045351 Free PMC article.
-
Vitamin D and thyroid disease: to D or not to D?Eur J Clin Nutr. 2015 Mar;69(3):291-6. doi: 10.1038/ejcn.2014.265. Epub 2014 Dec 17. Eur J Clin Nutr. 2015. PMID: 25514898 Review.
-
Improving the measurement of 25-hydroxyvitamin D.Clin Biochem Rev. 2005 Feb;26(1):33-6. Clin Biochem Rev. 2005. PMID: 16278775 Free PMC article. No abstract available.
-
Measurement of circulating 25-hydroxyvitamin D: A historical review.Pract Lab Med. 2015 May 12;2:1-14. doi: 10.1016/j.plabm.2015.04.001. eCollection 2015 Aug 1. Pract Lab Med. 2015. PMID: 28932799 Free PMC article. Review.