Abstract
Objective: To evaluate the performance of DNA amplification-based tests for the diagnosis of urethral chlamydia infection from the urine of asymptomatic young men.
Design: First-void urine was analyzed by two amplified DNA technologies, the ligase chain reaction (LCR), the polymerase chain reaction (PCR), and enzyme immunoassay (EIA). Specimens yielding discrepant results were subjected to retesting using either the original or a newly processed sample, and for evaluation of truly infected persons they were analyzed by the direct fluorescence antibody assay and by a second LCR directed against a segment of the gene encoding for the major outer membrane protein of Chlamydia trachomatis.
Setting: The military hospital in which military recruits underwent medical examination before departing for a United Nations mission.
Study group: Asymptomatic military recruits (705 young men) were screened between January and May 1994. In addition to providing urine specimens, the recruits completed questionnaires concerning previous genital infections and number of sexual partners.
Results: Overall prevalence of urethral chlamydia infection in the study group was 4.1% (29/705), as determined by confirmed results in all tests collectively. The performance of the DNA amplification methods was markedly better than that of the EIA antigen detection methods. Using an expanded gold standard, the sensitivity of the LCx assay was 93.1% (27/29) compared to 62.1% (18/29) by the PCR assay Amplicor and 37.9% (11/29) by EIA. Repeat testing after freezing and thawing increased the number of positive PCR results to equal the number of positive LCR results. There were three false-positive Amplicor results and no false-positive LCR results.
Conclusions: The LCx assay performed better than the Amplicor assay and appears reliable for urine testing. The low sensitivity of the Amplicor assay requires further evaluation of possible inhibitors of PCR in fresh specimens. It was found that freezing and thawing the specimens before testing enhanced the performance of PCR.
Similar articles
-
Comparison of performances of two commercially available tests, a PCR assay and a ligase chain reaction test, in detection of urogenital Chlamydia trachomatis infection.J Clin Microbiol. 1998 Jun;36(6):1489-93. doi: 10.1128/JCM.36.6.1489-1493.1998. J Clin Microbiol. 1998. PMID: 9620366 Free PMC article.
-
Detection of Chlamydia trachomatis by ligase chain reaction compared with polymerase chain reaction and cell culture in urogenital specimens.Genitourin Med. 1995 Dec;71(6):382-6. doi: 10.1136/sti.71.6.382. Genitourin Med. 1995. PMID: 8566978 Free PMC article.
-
Comparison of urine, first and second endourethral swabs for PCR based detection of genital Chlamydia trachomatis infection in male patients.Sex Transm Infect. 2001 Dec;77(6):423-6. doi: 10.1136/sti.77.6.423. Sex Transm Infect. 2001. PMID: 11714940 Free PMC article.
-
DFA, EIA, PCR, LCR and other technologies: what tests should be used for diagnosis of chlamydia infections?Immunol Invest. 1997 Jan-Feb;26(1-2):157-61. doi: 10.3109/08820139709048923. Immunol Invest. 1997. PMID: 9037620 Review.
-
DNA amplification assays: a new standard for diagnosis of Chlamydia trachomatis infections.Ann Acad Med Singap. 1995 Jul;24(4):627-33. Ann Acad Med Singap. 1995. PMID: 8849200 Review.
Cited by
-
Specimen processing and concentration of Chlamydia trachomatis added can influence false-negative rates in the LCx assay but not in the APTIMA Combo 2 assay when testing for inhibitors.J Clin Microbiol. 2003 Feb;41(2):778-82. doi: 10.1128/JCM.41.2.778-782.2003. J Clin Microbiol. 2003. PMID: 12574282 Free PMC article.
-
Ability of new APTIMA CT and APTIMA GC assays to detect Chlamydia trachomatis and Neisseria gonorrhoeae in male urine and urethral swabs.J Clin Microbiol. 2005 Jan;43(1):127-31. doi: 10.1128/JCM.43.1.127-131.2005. J Clin Microbiol. 2005. PMID: 15634960 Free PMC article.
-
Multicenter evaluation of the fully automated COBAS AMPLICOR PCR test for detection of Chlamydia trachomatis in urogenital specimens.J Clin Microbiol. 1999 Jan;37(1):74-80. doi: 10.1128/JCM.37.1.74-80.1999. J Clin Microbiol. 1999. PMID: 9854067 Free PMC article.
-
Emergency department screening for asymptomatic sexually transmitted infections.Am J Public Health. 2001 Mar;91(3):461-4. doi: 10.2105/ajph.91.3.461. Am J Public Health. 2001. PMID: 11236416 Free PMC article.
-
Ability of commercial ligase chain reaction and PCR assays to diagnose Chlamydia trachomatis infections in men by testing first-void urine.J Clin Microbiol. 1997 Apr;35(4):982-4. doi: 10.1128/jcm.35.4.982-984.1997. J Clin Microbiol. 1997. PMID: 9157168 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical