Comparative Study
. 1996 Jul-Aug;104(7-8):549-56.
doi: 10.1111/j.1699-0463.1996.tb04910.x.

Detection of penicillin resistance in Streptococcus pneumoniae by diffusion tests

Affiliations
Comparative Study

Detection of penicillin resistance in Streptococcus pneumoniae by diffusion tests

R L Poulsen et al. APMIS. 1996 Jul-Aug.

Abstract

Four different diffusion tests used to detect penicillin resistance in Streptococcus pneumoniae were evaluated for 34 penicillin-susceptible pneumococci (MIC < 0.1 microgram/ml), 35 intermediate pencillin-resistant (MIC 0.1-1.0 microgram/ml) and 23 penicillin-resistant strains (MIC > 2 micrograms/ml). The 1 microgram oxacillin disk from AB Biodisk, the 1 microgram oxacillin Neo-Sensitabs from Rosco, the 5 micrograms penicillin Low Neo-Sensitabs and the penicillin E test were tested on Mueller-Hinton blood agar, PDM Antibiotic Sensitivity Medium II supplemented with blood, and Danish Blood Agar. MICs obtained by the agar dilution method were used as reference. The 1 microgram oxacillin AB Biodisk was able to separate all the penicillin-susceptible pneumococci correctly from those with reduced penicillin susceptibility (MIC > or = 0.1 microgram/ml), whereas use of the 1 microgram oxacillin Neo-Sensitabs resulted in high frequencies (14-29%) of intermediate penicillin-resistant strains interpreted as penicillin susceptible. The 5 micrograms penicillin Low Neo-Sensitabs proved completely useless for detecting penicillin resistance in pneumococci. High rates of agreement (82-93%) were found between the penicillin E test and the reference MIC determination method on all the tested media.