Multicenter comparison of three commercial methods for quantification of human immunodeficiency virus type 1 RNA in plasma
- PMID: 8940441
- PMCID: PMC229452
- DOI: 10.1128/jcm.34.12.3016-3022.1996
Multicenter comparison of three commercial methods for quantification of human immunodeficiency virus type 1 RNA in plasma
Abstract
Three procedures for the quantification of human immunodeficiency virus type 1 (HIV-1) RNA from plasma were compared at three laboratories. The comparison involved the Quantiplex branched DNA assay (version 1.0) by Chiron Diagnostics, the NASBA-QT assay by Organon Teknika, and the Amplicor Monitor assay by Roche Molecular Systems. The laboratories performed each of the three assays with the same sets of reconstructed HIV-1-infected human plasma samples, cross-sectionally collected clinical plasma samples and longitudinally collected plasma samples from patients starting zidovudine therapy. Analysis of the reconstruction panel results for interlaboratory variation demonstrated that no laboratory differences in results were detected for any of the assays. A comparison of the reproducibilities of duplicate samples analyzed by batch and in separate assay runs demonstrated that the reproducibilities of the test results were similar within one assay and appeared to be independent of the HIV-1 concentration. The best reproducibility was obtained with the Quantiplex assay, but all three assays demonstrated equal reliability, which was independent of batched or unbatched analysis of replicate samples. Differences in the absolute concentrations calculated were observed for the assays, in particular in the analysis of reconstructed samples. In all assays, similar changes in plasma HIV-1 RNA concentrations were determined for longitudinally collected clinical samples.
Similar articles
-
Comparison of the QUANTIPLEX HIV-1 RNA 2.0 assay with the AMPLICOR HIV-1 MONITOR 1.0 assay for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma of patients receiving stavudine-didanosine combination therapy.J Clin Microbiol. 1998 Nov;36(11):3392-5. doi: 10.1128/JCM.36.11.3392-3395.1998. J Clin Microbiol. 1998. PMID: 9774604 Free PMC article.
-
Comparative evaluation of NASBA HIV-1 RNA QT, AMPLICOR-HIV monitor, and QUANTIPLEX HIV RNA assay, three methods for quantification of human immunodeficiency virus type 1 RNA in plasma.J Clin Microbiol. 1996 May;34(5):1058-64. doi: 10.1128/jcm.34.5.1058-1064.1996. J Clin Microbiol. 1996. PMID: 8727875 Free PMC article.
-
Multicenter comparison of Roche COBAS AMPLICOR MONITOR version 1.5, Organon Teknika NucliSens QT with Extractor, and Bayer Quantiplex version 3.0 for quantification of human immunodeficiency virus type 1 RNA in plasma.J Clin Microbiol. 2000 Nov;38(11):4034-41. doi: 10.1128/JCM.38.11.4034-4041.2000. J Clin Microbiol. 2000. PMID: 11060065 Free PMC article.
-
Comparison of the Quantiplex version 3.0 assay and a sensitized Amplicor monitor assay for measurement of human immunodeficiency virus type 1 RNA levels in plasma samples.J Clin Microbiol. 1999 Nov;37(11):3612-4. doi: 10.1128/JCM.37.11.3612-3614.1999. J Clin Microbiol. 1999. PMID: 10523562 Free PMC article.
-
Molecular-based methods for quantifying HIV viral load.AIDS Patient Care STDS. 2004 Feb;18(2):75-9. doi: 10.1089/108729104322802506. AIDS Patient Care STDS. 2004. PMID: 15006182 Review.
Cited by
-
The role of the virology laboratory in the management of hepatitis C virus infection.J Clin Virol. 2002 Jul;25(1):3-13. doi: 10.1016/s1386-6532(02)00026-4. J Clin Virol. 2002. PMID: 12126716 Free PMC article. Review.
-
Comparison of the Amplicor HIV-1 monitor test and the nucleic acid sequence-based amplification assay for quantitation of human immunodeficiency virus RNA in plasma, serum, and plasma subjected to freeze-thaw cycles.J Clin Microbiol. 1997 Dec;35(12):3288-91. doi: 10.1128/jcm.35.12.3288-3291.1997. J Clin Microbiol. 1997. PMID: 9399536 Free PMC article.
-
Comparison of the QUANTIPLEX HIV-1 RNA 2.0 assay with the AMPLICOR HIV-1 MONITOR 1.0 assay for quantitation of levels of human immunodeficiency virus type 1 RNA in plasma of patients receiving stavudine-didanosine combination therapy.J Clin Microbiol. 1998 Nov;36(11):3392-5. doi: 10.1128/JCM.36.11.3392-3395.1998. J Clin Microbiol. 1998. PMID: 9774604 Free PMC article.
-
Determinations of levels of human immunodeficiency virus type 1 RNA in plasma: reassessment of parameters affecting assay outcome. TUBE Meeting Workshop Attendees. Technology Utilization for HIV-1 Blood Evaluation and Standardization in Pediatrics.J Clin Microbiol. 1998 Jun;36(6):1471-9. doi: 10.1128/JCM.36.6.1471-1479.1998. J Clin Microbiol. 1998. PMID: 9620364 Free PMC article. Review. No abstract available.
-
Current practices in laboratory monitoring of HIV infection.Indian J Med Res. 2011 Dec;134(6):801-22. doi: 10.4103/0971-5916.92627. Indian J Med Res. 2011. PMID: 22310815 Free PMC article. Review.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources