Local delivery of heparin with balloon angioplasty: results of a prospective randomized trial
- PMID: 8951752
- DOI: 10.1016/s1051-0443(96)70860-6
Local delivery of heparin with balloon angioplasty: results of a prospective randomized trial
Abstract
Purpose: To evaluate the effectiveness of local delivery of heparin via hydrogel-coated balloons in the treatment of vascular stenoses associated with hemodialysis access.
Materials and methods: This was a randomized, prospective trial comparing treatment with hydrogel-coated balloon catheters delivered with heparin coating (n = 33) and without (n = 26). All patients were undergoing hemodialysis, and all stenoses involved the venous anastomosis of a dialysis graft or a native vein. The heparin-treated balloons were soaked in concentrated heparin and delivered in a protected manner to help prevent washout of heparin.
Results: The mean primary patencies were 143 days with heparin treatment and 214 days without heparin (P = .174). The mean assisted primary patencies were 165 days with heparin and 194 days without (P = .315). The mean secondary patencies were 351 days with heparin and 384 without (P = .81).
Conclusion: In this population with this technique, the treatment outcome of venous outflow stenosis in patients with dialysis grafts is not improved with local delivery of heparin.