Gradual reduction of endotracheal tube diameter during mechanical ventilation via different humidification devices
- PMID: 8968181
- DOI: 10.1097/00000542-199612000-00015
Gradual reduction of endotracheal tube diameter during mechanical ventilation via different humidification devices
Abstract
Background: Limited data suggest that increased resistance to flow within endotracheal tubes (ETT) may occur in patients whose lungs are mechanically ventilated for more than 48 h, especially when airway humidification is inadequate. This could lead to sudden ETT obstruction or induce excessive loading during spontaneous breathing.
Methods: Twenty-three such patients were randomly assigned to three types of airway humidifier based on three different working principles: a Fisher Paykell hot water system (n = 7), a Pall BB2215 heat and moisture exchanger (HME) hydrophobic filter (n = 8), and a Dar Hygrobac 35254111 HME hygroscopic filter (n = 8). The decrease in internal pressure along the ETT and the flow rate were measured in each patient every 2 days. An "effective inner diameter" was derived from these measurements and allowed the inner ETT configuration to be monitored.
Results: On the first day of intubation, the mean diameter was similar in the three groups, and was slightly smaller than the in vitro diameter (mean +/- SD: 7.6 +/- 0.6 mm for Fisher-Paykell, 7.7 +/- 0.4 for Pall, and 7.5 +/- 0.4 for Dar). The mean diameter tended to decrease from day to day. At the end of the study, the overall reduction in mean diameter was significantly greater with the hydrophobic HME (Pall) than with the two other systems (Pall: -6.5 +/- 4% vs. 2.5 +/- 2.5% for Dar and 1.5 +/- 3% for Fisher-Paykell; P < 0.01 with analysis of variance). The same was true of the mean reduction in effective inner ETT diameter expressed per day of ventilation (-1.6 +/- 1.5% per day for Pall vs. -0.5 +/- 0.4% for Dar and -0.2 +/- 0.4% for Fisher-Paykell; P < 0.01). In four patients, the ETT became obstructed and emergency repeated tracheal intubation was required. The Pall HME and the Fisher-Paykell system were being used in three and one patient, respectively. Before obstruction, the reduction in ETT diameter was significantly greater for these four patients than for the remaining 23 patients (7.8 +/- 1.4% vs. 3.1 +/- 4.1%; P < 0.01).
Conclusions: During prolonged mechanical ventilation, significant alterations in inner ETT configuration occur frequently and are influenced by the type of humidification device used. In vivo monitoring of ETT mechanical properties might be clinically useful.
Similar articles
-
Bedside evaluation of efficient airway humidification during mechanical ventilation of the critically ill.Chest. 1999 Jun;115(6):1646-52. doi: 10.1378/chest.115.6.1646. Chest. 1999. PMID: 10378563 Clinical Trial.
-
Long-term effects of different humidification systems on endotracheal tube patency: evaluation by the acoustic reflection method.Anesthesiology. 2004 Apr;100(4):782-8. doi: 10.1097/00000542-200404000-00006. Anesthesiology. 2004. PMID: 15087611
-
Detection of endotracheal tube obstruction by analysis of the expiratory flow signal.Intensive Care Med. 1998 Nov;24(11):1163-72. doi: 10.1007/s001340050740. Intensive Care Med. 1998. PMID: 9876979
-
Devices used to humidify respired gases.Respir Care Clin N Am. 2006 Jun;12(2):165-82. doi: 10.1016/j.rcc.2006.03.012. Respir Care Clin N Am. 2006. PMID: 16828689 Review.
-
[Artificial humidification of inspired gas--status of knowledge and technique].Anaesthesiol Reanim. 1997;22(6):153-8. Anaesthesiol Reanim. 1997. PMID: 9487786 Review. German.
Cited by
-
Internally coated endotracheal tubes with silver sulfadiazine in polyurethane to prevent bacterial colonization: a clinical trial.Intensive Care Med. 2008 Jun;34(6):1030-7. doi: 10.1007/s00134-008-1100-1. Epub 2008 Apr 17. Intensive Care Med. 2008. PMID: 18418571 Clinical Trial.
-
Towards Zero Endotracheal Tube Occlusion: A Reevaluation of Passive Humidifiers Testing and Selection.Am J Respir Crit Care Med. 2025 Feb;211(2):154-156. doi: 10.1164/rccm.202410-1889VP. Am J Respir Crit Care Med. 2025. PMID: 39637367 Free PMC article. No abstract available.
-
Heat and moisture exchangers versus heated humidifiers for mechanically ventilated adults and children.Cochrane Database Syst Rev. 2017 Sep 14;9(9):CD004711. doi: 10.1002/14651858.CD004711.pub3. Cochrane Database Syst Rev. 2017. PMID: 28905374 Free PMC article.
-
Endotracheal tube intraluminal diameter narrowing after mechanical ventilation: use of acoustic reflectometry.Intensive Care Med. 2004 Dec;30(12):2204-9. doi: 10.1007/s00134-004-2465-4. Epub 2004 Oct 2. Intensive Care Med. 2004. PMID: 15580475
-
Nebulized antibiotics in mechanically ventilated patients: a challenge for translational research from technology to clinical care.Ann Intensive Care. 2017 Dec;7(1):78. doi: 10.1186/s13613-017-0301-6. Epub 2017 Aug 1. Ann Intensive Care. 2017. PMID: 28766281 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical