A simple technique for creating hybrid fixators using a modified AO single adjustable clamp
- PMID: 8990036
- DOI: 10.1097/00005131-199701000-00016
A simple technique for creating hybrid fixators using a modified AO single adjustable clamp
Abstract
Hybrid external fixators with limited internal fixation have become popular in the treatment of complex intraarticular fractures of the tibia. A wide variety of hybrid external fixation systems are commercially available, but these frames are expensive and not always readily available. The authors present a simple hybrid frame constructed from the AO tubular and Ilizarov type external fixator sets using a modified AO single adjustable clamp. This hybrid frame is easy to apply, versatile, and significantly less expensive than other commercially available adaptors and frames.