[The relation between the density of blood and the arterial pressure in animal experiments and in patients during hemodialysis (author's transl)]
- PMID: 899140
[The relation between the density of blood and the arterial pressure in animal experiments and in patients during hemodialysis (author's transl)]
Abstract
The density of arterial blood was measured continuously in animals and in human beings with an accuracy of 10(-5) g/ml by the application of the so-called mechanical oscillator technique. The density measuring device is inserted into an arterio-venous shunt. The continuous reading of the density of arterial blood allows to observe the effect of fluid shifts between the interstitial space and the plasma space. For example, lowering of the capillary pressure by the constriction of the precapillary resistance decreases the blood density by the inflow of interstitial fluid. Strong vasoconstriction apparently increases the capillary pressure and thus raises the blood density by the loss of plasma-ultrafiltrate. During hemodialysis the fluid loss into the dialysator tends to increase the blood density. However, the regulatory mechanisms of the circulation are able to keep the density of the blood at a constant value as long as possible. An increase of the density indicates a decrease of the blood volume. The new method thus allows a continuous control of the course of hemodialysis, and to avoid a critical decrease of the blood volume and of the blood pressure.
Similar articles
-
[Continuous determination of the blood density during haemodialysis (author's transl)].Wien Klin Wochenschr. 1979 Nov 23;91(22):762-5. Wien Klin Wochenschr. 1979. PMID: 516747 German.
-
New blood volume monitoring method for hemodialysis: A-V pressure gradient measurement by synchronized one-point reading.Artif Organs. 2004 Jul;28(7):683-9. doi: 10.1111/j.1525-1594.2004.07302.x. Artif Organs. 2004. PMID: 15209863
-
Distribution of body fluids: local mechanisms guarding interstitial fluid volume.J Physiol (Paris). 1984;79(6):395-400. J Physiol (Paris). 1984. PMID: 6399307 Review.
-
Device and technique for extracorporeal blood volume sequestration during hemodialysis.ASAIO J. 2006 Nov-Dec;52(6):662-9. doi: 10.1097/01.mat.0000235812.55055.5b. ASAIO J. 2006. PMID: 17117056 Clinical Trial.
-
[Automatic system control and its application in dialysis].G Ital Nefrol. 2007 Nov-Dec;24 Suppl 40:s69-79. G Ital Nefrol. 2007. PMID: 18034415 Review. Italian.
Cited by
-
[Application of the mechanical oscillator technique for the measurement of blood density and hematocrit (author's transl)].Klin Wochenschr. 1979 Nov 2;57(21):1163-7. doi: 10.1007/BF01491756. Klin Wochenschr. 1979. PMID: 513608 German.
-
Continuous measurement of blood volume using light reflection: method and validation.Med Biol Eng Comput. 1993 Jul;31(4):412-5. doi: 10.1007/BF02446697. Med Biol Eng Comput. 1993. PMID: 8231306 No abstract available.
-
Closed Loop Ultrafiltration Feedback Control in Hemodialysis: A Narrative Review.Toxins (Basel). 2024 Aug 10;16(8):351. doi: 10.3390/toxins16080351. Toxins (Basel). 2024. PMID: 39195761 Free PMC article. Review.
-
The measurement of blood density and its meaning.Basic Res Cardiol. 1989 Mar-Apr;84(2):111-24. doi: 10.1007/BF01907921. Basic Res Cardiol. 1989. PMID: 2658951 Review.
-
Transvascular fluid shifts accompanying short bodily activity in the alert rat and their relation to arterial pressure.Basic Res Cardiol. 1982 Jul-Aug;77(4):431-48. doi: 10.1007/BF02005343. Basic Res Cardiol. 1982. PMID: 7150211 No abstract available.