Parameters that influence accuracy and precision of quantitative coronary arteriography
- PMID: 8993990
- DOI: 10.1007/BF01797741
Parameters that influence accuracy and precision of quantitative coronary arteriography
Abstract
The limited resolution of any imaging system causes edge blurring of objects acquired with X-ray. In digital angiography, this effect combined with noise gives rise to systematic and random errors in the determination of vessel dimensions. The influence of bandwidth limitation on the estimation of tube diameter is established by a theoretical approach: it leads to over and underestimations of catheter and vessel diameter dimensions. Therefore a correction is proposed that counterbalances the point spread function (PSF) offset. The residual inaccuracy and the variability of measurement of phantom tubes are analyzed and evaluated in controlled conditions. Some of the parameters which govern their extent are identified: field-of-view, catheter size, concentration of contrast agent.
Similar articles
-
Quantitative coronary arteriography: efficient correction of catheter calibrated vessel measurement.Medinfo. 1995;8 Pt 1:724. Medinfo. 1995. PMID: 8591310
-
Inaccuracy of quantitative coronary arteriography when analyzed from S-VHS videotape.Cathet Cardiovasc Diagn. 1996 Jan;37(1):32-8. doi: 10.1002/(SICI)1097-0304(199601)37:1<32::AID-CCD8>3.0.CO;2-5. Cathet Cardiovasc Diagn. 1996. PMID: 8770476
-
[Effect of sharpening filter on vessel diameter measured by quantitative coronary arteriography].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2002 May;58(5):694-9. doi: 10.6009/jjrt.kj00001364429. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2002. PMID: 12520240 Japanese.
-
Adaptive approach to accurate analysis of small-diameter vessels in cineangiograms.IEEE Trans Med Imaging. 1997 Feb;16(1):87-95. doi: 10.1109/42.552058. IEEE Trans Med Imaging. 1997. PMID: 9050411
-
The influence of image enhancement and reconstruction on quantitative coronary arteriography.Int J Card Imaging. 1995 Dec;11(4):211-21. doi: 10.1007/BF01145189. Int J Card Imaging. 1995. PMID: 8596059
Cited by
-
Calibration-free coronary artery measurements for interventional device sizing using inverse geometry x-ray fluoroscopy: in vivo validation.J Med Imaging (Bellingham). 2014 Oct;1(3):033504. doi: 10.1117/1.JMI.1.3.033504. J Med Imaging (Bellingham). 2014. PMID: 25544948 Free PMC article.
-
The 6sigma quality approach for quantitative arteriography performance improvement.Int J Cardiovasc Imaging. 2002 Apr;18(2):77-92. doi: 10.1023/a:1014658913126. Int J Cardiovasc Imaging. 2002. PMID: 12108912
-
The Anatomic-Functional Duality of So-called 'Significant' Atherosclerotic Stenosis - Update on Invasive Diagnostic Strategies in Coronaropathy.Interv Cardiol. 2013 Aug;8(2):112-117. doi: 10.15420/icr.2013.8.2.112. Interv Cardiol. 2013. PMID: 29588762 Free PMC article.
References
MeSH terms
LinkOut - more resources
Other Literature Sources