Local intraarterial thrombolysis: in vitro comparison between automatic and manual pulse-spray infusion

Affiliations
Comparative Study

Local intraarterial thrombolysis: in vitro comparison between automatic and manual pulse-spray infusion

J J Froelich et al. Cardiovasc Intervent Radiol. 1996 Nov-Dec.

Abstract

Purpose: Manual and automatic pulse-spray infusion techniques are compared in vitro to evaluate the efficacy of thrombolysis and the distribution of urokinase and saline solution within thrombus using a pulse-spray catheter.

Methods: A pulse-spray catheter was introduced into a human thrombus within a stenotic flow model. Automatic and manual pulsed infusion of urokinase and automatic pulsed infusion of saline solution were compared. To quantify the efficacy of thrombolysis, pressure gradients were recorded proximal and distal to the thrombus and during the course of infusion. Distribution of infused urokinase was assessed radiographically.

Results: The fastest and most homogeneous dissolution of the thrombus was achieved with automatic pulsed infusion of urokinase, shown by decreasing transthrombotic pressure gradients (p < 0.001, Wilcoxon, matched pairs). Manual pulsed infusion of urokinase or saline solution resulted in inhomogeneous thrombus dissolution and delayed thrombolysis.

Conclusion: Application of automatic pulse-spray injectors seems beneficial for more effective and homogeneous intraarterial pulse-spray thrombolysis when compared with conventional manual pulsed technique.

PubMed Disclaimer

Similar articles

References

    1. Radiology. 1983 Sep;148(3):677-82 - PubMed
    1. AJR Am J Roentgenol. 1986 Sep;147(3):621-6 - PubMed
    1. Radiology. 1990 Mar;174(3 Pt 2):1056-7 - PubMed
    1. N Engl J Med. 1982 Dec 23;307(26):1627-30 - PubMed
    1. Cardiovasc Intervent Radiol. 1991 Nov-Dec;14 (6):334-7 - PubMed

Publication types

MeSH terms

Substances

LinkOut - more resources