Comparative Study
. 1997 Jan;12(1):25-31.
doi: 10.1016/s0883-5403(97)90043-7.

Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups

Affiliations
Comparative Study

Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups

V G Williams 2nd et al. J Arthroplasty. 1997 Jan.

Abstract

Locking mechanisms and metal-liner interface surfaces of six modular acetabular systems were evaluated to determine their effect on micromotion and backside wear of the polyethylene liner. Rotational and axial motion between the metal shell and polyethylene liner was measured in the Duraloc (DePuy, Warsaw, IN), Harris-Galante (Zimmer, Warsaw, IN), Impact (Biomet, Warsaw, IN), Lip Loc (Biomet), Precision Osteoloc (Howmedica, Rutherford, NJ), and Reflection (Smith & Nephew Orthopaedics, Memphis, TN) designs at the start of each test, and at 1 million, 5 million, and 10 million cycles. At 10 million cycles, the Lip Loc and Reflection cups had significantly lower rim micromotion than the Duraloc and Harris-Galante cups (F < .0010). The Impact, Precision Osteoloc, and Reflection cups had significantly lower rim subsidence than the Harris-Galante cup (F < .0025). The Harris-Galante cup had significantly greater rotational micromotion than the Lip Loc cup (F < .0074), and had significantly greater interface slippage than the Impact and Reflection cups (F < .0070). The Lip Loc produced significantly lower dome micromotion than the Harris-Galante (F < .0300). The Lip Loc and Reflection cups had significantly less backside wear than the Duraloc and Harris-Galante cups (P < .0001), the Impact cup (P < .0243), and the Precision Osteoloc (P < .0027) cup.

PubMed Disclaimer