Enzyme-linked immunosorbent assay for the detection of total cathepsin H in human tissue cytosols and sera
- PMID: 9050938
- DOI: 10.1016/s0022-1759(96)00218-9
Enzyme-linked immunosorbent assay for the detection of total cathepsin H in human tissue cytosols and sera
Abstract
An enzyme-linked immunosorbent assay (ELISA) was constructed for the determination of total human cathepsin H concentration in clinical samples. Utilising monoclonal and polyclonal antibodies, raised to human liver cathepsin H, the assay is able to detect a mature protein, a precursor molecule and enzyme-inhibitor complexes. The test system permits sensitive and reliable detection of analyte either in tissue cytosols or in sera. The detection limit is 2 ng/ml (n = 10, mean of zero standard +/- 3 SD). The average recovery of cathepsin H, added to the low content samples, was 95.3% +/- 1.8%. The within-run and between-run coefficient of variance (CV) varied from 2.3% to 8.9% and 12.7% to 16.4%, respectively, indicating satisfactory reproducibility of the method. The level of cathepsin H was defined in tissue cytosols of human heart, muscle and kidney and in sera from 30 healthy individuals. Additionally, cathepsin H was measured in sera from 55 patients with primary skin melanoma and from 42 patients with metastatic melanoma. The mean cathepsin H level was significantly higher for both groups of patients compared to normal sera level, being highest for metastatic melanoma patients.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Molecular Biology Databases