Method to produce American Thoracic Society flow-time waveforms using a mechanical pump
- PMID: 9073007
Method to produce American Thoracic Society flow-time waveforms using a mechanical pump
Abstract
The American Thoracic Society (ATS) recently adopted a new set of 26 standard flow-time waveforms for use in testing both diagnostic and monitoring devices. Some of these waveforms have a higher frequency content than present in the ATS-24 standard volume-time waveforms, which, when produced by a mechanical pump, may result in a pump flow output that is less than the desired flow due to gas compression losses within the pump. To investigate the effects of gas compression, a mechanical pump was used to generate the necessary flows to test mini-Wright and Assess peak expiratory flow (PEF) meters. Flow output from the pump was measured by two different independent methods, a pneumotachometer and a method based on piston displacement and pressure measured within the pump. Measuring output flow based on piston displacement and pressure has been validated using a pneumotachometer and mini-Wright PEF meter, and found to accurately measure pump output. This method introduces less resistance (lower back-pressure) and dead space volume than using a pneumotachometer in series with the meter under test. Pump output flow was found to be lower than the desired flow both with the mini-Wright and Assess meters (for waveform No. 26, PEFs 7.1 and 10.9% lower, respectively). To compensate for losses due to gas compression, we have developed a method of deriving new input waveforms, which, when used to drive a commercially available mechanical pump, accurately and reliably produces the 26 ATS flow-time waveforms, even those with the fastest rise-times.
Similar articles
-
Gas compression artefacts when testing peak expiratory flow meters with mechanically-driven syringes.Eur Respir J. 1997 Apr;10(4):901-4. Eur Respir J. 1997. PMID: 9150332
-
Standard flow-time waveforms for testing of PEF meters.Am J Respir Crit Care Med. 1995 Aug;152(2):696-701. doi: 10.1164/ajrccm.152.2.7633728. Am J Respir Crit Care Med. 1995. PMID: 7633728
-
Nonlinear model for offline correction of pulmonary waveform generators.IEEE Trans Biomed Eng. 2002 Dec;49(12 Pt 2):1567-73. doi: 10.1109/TBME.2002.805484. IEEE Trans Biomed Eng. 2002. PMID: 12549738
-
Beyond the peak flow meter: newer technologies for determining and documenting changes in lung function in the workplace.Occup Med. 2000 Apr-Jun;15(2):411-20. Occup Med. 2000. PMID: 10769346 Review.
-
Instrumentation for spirometry.Occup Med. 1993 Apr-Jun;8(2):397-407. Occup Med. 1993. PMID: 8506514 Review.
Cited by
-
Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device.Thorax. 2003 May;58(5):411-6. doi: 10.1136/thorax.58.5.411. Thorax. 2003. PMID: 12728161 Free PMC article.
-
The syringe potentiometer: a low-cost device for pneumotachograph calibration.J Appl Physiol (1985). 2019 Oct 1;127(4):1150-1162. doi: 10.1152/japplphysiol.00150.2019. Epub 2019 Sep 5. J Appl Physiol (1985). 2019. PMID: 31487222 Free PMC article.
-
A Smart Phone Based Handheld Wireless Spirometer with Functions and Precision Comparable to Laboratory Spirometers.Sensors (Basel). 2019 May 31;19(11):2487. doi: 10.3390/s19112487. Sensors (Basel). 2019. PMID: 31159155 Free PMC article.
-
A Fast Calibration Method for Pneumotachograph with a 3L Syringe.Bioengineering (Basel). 2023 Sep 7;10(9):1053. doi: 10.3390/bioengineering10091053. Bioengineering (Basel). 2023. PMID: 37760155 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources