. 1997 Mar;117(2):274-7.
doi: 10.3109/00016489709117786.

Cochlear implant deep electrode insertion: extent of insertional trauma

Affiliations

Cochlear implant deep electrode insertion: extent of insertional trauma

W Gstoettner et al. Acta Otolaryngol. 1997 Mar.

Abstract

We have recently undertaken deep insertions of the Combi-40 cochlear implant electrode (Med-E1 Corp., Innsbruck, Austria) into apical regions of the scala tympani using a cochleostomy approach. In order to examine the extent of the insertional trauma, 12 fresh human temporal bones were implanted with original Combi-40 electrodes. The specimens were histologically processed with the implants in place by employing a sawing and grinding technique. In most cases, only very discrete distortions of the epithelium of the spiral ligament occurred within the middle cochlear turns. Furthermore, a slight displacement of the basilar membrane caused by the electrode was occasionally seen. However, in 2 cases more severe damage such as basilar membrane rupture and electrode displacement was found. Attempts to insert the electrode beyond the point of first resistance resulted in electrode kinking within the basal cochlear turn with subsequent fracture of the osseous spiral lamina. According to our results, deep electrode insertions do not aggravate the insertional trauma provided no force is applied when resistance is felt.

PubMed Disclaimer

LinkOut - more resources