A novel method for nonfluoroscopic catheter-based electroanatomical mapping of the heart. In vitro and in vivo accuracy results
- PMID: 9118532
- DOI: 10.1161/01.cir.95.6.1611
A novel method for nonfluoroscopic catheter-based electroanatomical mapping of the heart. In vitro and in vivo accuracy results
Abstract
Background: Cardiac mapping is essential for understanding the mechanisms of arrhythmias and for directing curative procedures. A major limitation of the current methods is the inability to accurately relate local electrograms to their spatial orientation. The objective of this study was to present and test the accuracy of a new method for nonfluoroscopic, catheter-based, endocardial mapping.
Methods and results: The method is based on using a new locatable catheter connected to an endocardial mapping and navigating system. The system uses magnetic technology to accurately determine the location and orientation of the catheter and simultaneously records the local electrogram from its tip. By sampling a plurality of endocardial sites, the system reconstructs the three-dimensional geometry of the chamber, with the electrophysiological information color-coded and superimposed on the anatomy. The accuracy of the system was tested in both in vitro and in vivo studies and was found to be highly reproducible (SD, 0.16 +/- 0.02 [mean +/- SEM] and 0.74 +/- 0.13 mm) and accurate (mean errors, 0.42 +/- 0.05 and 0.73 +/- 0.03 mm). In further studies, electroanatomical mapping of the cardiac chambers was performed in 34 pigs. Both the geometry and activation sequence were repeatable in all pigs.
Conclusions: The new mapping method is highly accurate and reproducible. The ability to combine electrophysiological and spatial information provides a unique tool for both research and clinical electrophysiology. Consequently, the main shortcomings of conventional mapping-namely, prolonged x-ray exposure, low spatial resolution, and the inability to accurately navigate to a predefined site-can all be overcome with this new method.
Similar articles
-
3D cardiac imaging of electromechanical coupling.Adv Exp Med Biol. 1997;430:303-11. doi: 10.1007/978-1-4615-5959-7_26. Adv Exp Med Biol. 1997. PMID: 9330739
-
Hemodynamic evaluation of the heart with a nonfluoroscopic electromechanical mapping technique.Circulation. 1997 Nov 18;96(10):3672-80. doi: 10.1161/01.cir.96.10.3672. Circulation. 1997. PMID: 9396470
-
Use of novel nonfluoroscopic three-dimensional electroanatomic mapping system to monitor and analyze heart surgery in animal models.Chest. 2004 May;125(5):1830-6. doi: 10.1378/chest.125.5.1830. Chest. 2004. PMID: 15136397
-
Non-fluoroscopic catheter-based mapping systems in cardiac electrophysiology--from approved clinical indications to novel research usage.Int J Med Robot. 2006 Mar;2(1):21-7. doi: 10.1002/rcs.65. Int J Med Robot. 2006. PMID: 17520610 Review.
-
Recent advances in cardiac mapping techniques.Curr Cardiol Rep. 1999 Jul;1(2):149-56. doi: 10.1007/s11886-999-0074-0. Curr Cardiol Rep. 1999. PMID: 10980835 Review.
Cited by
-
First in vivo use of a capacitive micromachined ultrasound transducer array-based imaging and ablation catheter.J Ultrasound Med. 2012 Feb;31(2):247-56. doi: 10.7863/jum.2012.31.2.247. J Ultrasound Med. 2012. PMID: 22298868 Free PMC article.
-
Noninvasive mapping of transmural potentials during activation in swine hearts from body surface electrocardiograms.IEEE Trans Med Imaging. 2012 Sep;31(9):1777-85. doi: 10.1109/TMI.2012.2202914. Epub 2012 Jun 6. IEEE Trans Med Imaging. 2012. PMID: 22692900 Free PMC article.
-
Transendocardial cell injection is not superior to intracoronary infusion in a porcine model of ischaemic cardiomyopathy: a study on delivery efficiency.J Cell Mol Med. 2012 Nov;16(11):2768-76. doi: 10.1111/j.1582-4934.2012.01594.x. J Cell Mol Med. 2012. PMID: 22697212 Free PMC article.
-
Efficacy of electroanatomic mapping in the catheter ablation of premature ventricular contractions originating from the right ventricular outflow tract.J Interv Card Electrophysiol. 2007 Sep;19(3):187-94. doi: 10.1007/s10840-007-9160-7. Epub 2007 Sep 22. J Interv Card Electrophysiol. 2007. PMID: 17891452 Clinical Trial.
-
The importance of right atrial pacing electrode position and pacing configuration for intra-atrial and inter-atrial conduction times.J Interv Card Electrophysiol. 2000 Jun;4(2):405-13. doi: 10.1023/a:1009858601764. J Interv Card Electrophysiol. 2000. PMID: 10936006
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical