Poor clinical performance of the Wessex porcine heart valve bioprosthesis at nine years' follow up
- PMID: 9155609
- PMCID: PMC484724
- DOI: 10.1136/hrt.77.4.319
Poor clinical performance of the Wessex porcine heart valve bioprosthesis at nine years' follow up
Abstract
Objective: To assess the long term performance of the Wessex porcine bioprostheses implanted in a consecutive series of patients.
Design: A retrospective case series.
Patients: Between January 1985 and July 1991, 184 Wessex bioprostheses (78 mitral, 102 aortic, and 4 tricuspid) were implanted in 150 patients. The patients were 55% (83/150) male and 45% (67/150) female; mean age was 60 (SD 10) years.
Results: Hospital mortality was 9.3% (14/150). Total follow up was 696 patient-years (mean 4.7 years per patient). Linearised rates (events per 100 patient-years (SEM) for postoperative complications for patients with isolated mitral valve replacement, isolated aortic valve replacement, and multiple valve replacement were, respectively: late mortality: 4.7 (1.6), 3.3 (0.9), and 4.9 (1.9); thromboembolism: 5.8 (1.8), 3.0 (0.9), and 2.8 (1.4); valve thrombosis: 1.0 (0.7), 0.3 (0.3), and 0.7 (0.7); structural failure: 5.8 (1.7), 1.9 (0.7), and 7.1 (2.2). Actuarial freedom from complications at nine years (70% confidence interval) was: late mortality: 61 (9)%, 57 (13)%, and 59 (12)%; thromboembolism and valve thrombosis: 71 (9)%, 79 (6)%, and 81 (8)%; structural failure: 33 (14)%, 50 (16)%, and 12 (14)%; all valve related morbidity/mortality: 31 (10)%, 21 (11)%, and 7 (9)%. Stent fractures appeared in 11 of 17 explanted prostheses; actuarial freedom from stent fracture at nine years was 66 (12)%.
Conclusions: The Wessex bioprosthesis is associated with high thrombogenicity, early structural dysfunction, and a high valve related morbidity/mortality which justifies very close follow up of patients fitted with them.
Similar articles
-
Stent fracture in Wessex porcine heart valve bioprostheses.Artif Organs. 1997 Feb;21(2):116-20. doi: 10.1111/j.1525-1594.1997.tb00347.x. Artif Organs. 1997. PMID: 9028493
-
Hancock II porcine bioprosthesis: excellent durability at intermediate-term follow-up.J Am Coll Cardiol. 1994 Sep;24(3):676-82. doi: 10.1016/0735-1097(94)90014-0. J Am Coll Cardiol. 1994. PMID: 8077538
-
An eight-year experience with porcine bioprosthetic cardiac valves.J Thorac Cardiovasc Surg. 1986 Jun;91(6):910-7. J Thorac Cardiovasc Surg. 1986. PMID: 3713240
-
Five-year evaluation of the Carpentier-Edwards porcine bioprosthesis.J Thorac Cardiovasc Surg. 1984 Sep;88(3):324-33. J Thorac Cardiovasc Surg. 1984. PMID: 6381888 Review.
-
Mitral valve disease: if the mitral valve is not reparable/failed repair, is bioprosthesis suitable for replacement?Eur J Cardiothorac Surg. 2009 Jan;35(1):104-10. doi: 10.1016/j.ejcts.2008.08.021. Epub 2008 Dec 3. Eur J Cardiothorac Surg. 2009. PMID: 19056294 Review.
Cited by
-
The United Kingdom Heart Valve Registry: the first 10 years.Heart. 1997 Apr;77(4):295-6. doi: 10.1136/hrt.77.4.295. Heart. 1997. PMID: 9155599 Free PMC article. No abstract available.
References
MeSH terms
LinkOut - more resources
Full Text Sources