[Shortening of the measurement time by 1H-MR turbo spectroscopic imaging of the brain]
- PMID: 9156593
- DOI: 10.1055/s-2007-1015413
[Shortening of the measurement time by 1H-MR turbo spectroscopic imaging of the brain]
Abstract
Purpose: Development of a new technique for reduction of measurement time in 1H-MR spectroscopic imaging of the brain. Optimisation of the sequence parameters in volunteer and in patient examinations and comparison to the results obtained with conventional 2 D-SI.
Methods: Examination of 20 healthy volunteers and 5 patients in a 1.5 T whole-body MR system. In "turbo-spectroscopic imaging" (TSI) sequences, a train of spin-echo signals with different phase encoding is acquired after each 90 degrees excitation. 32 x 32 matrix elements covered a field of view of 20 cm, and additional volume selection was performed by double spin echo excitation. Measurement duration 9 min with acquisition of four phase encoding steps per TR interval, whereas the corresponding 2 D-SI sequence (TR/TE 2000/272 ms) took 30 min.
Results: The TSI data sets yielded maps of the regional distribution of metabolite concentrations with a quality comparable to the 2 D-SI results. Signal homogeneity and delineation of brain lesions, however, were superior in conventional spectroscopic imaging. The T2 relaxation of the metabolites required a reduced sampling interval for each phase-encoded echo, and hence the frequency resolution of the corresponding TSI spectra was not always sufficient for separating choline and creatine signals.
Conclusion: With measurement durations < 10 min the TSI technique allows in clinical studies a combination with single-voxel MRS for accurate quantification and with a diagnostic MRI within a total examination time of less than one hour.
Similar articles
-
[1H-MR Spectroscopy of brain tumors in the course of radiation therapy: Use of fast spectroscopic imaging and single-voxel spectroscopy for diagnosing recurrence].Rofo. 2002 Jan;174(1):33-42. doi: 10.1055/s-2002-19541. Rofo. 2002. PMID: 11793282 German.
-
Selective maximization of (31)P MR spectroscopic signals of in vivo human brain metabolites at 3T.J Magn Reson Imaging. 2007 Mar;25(3):628-34. doi: 10.1002/jmri.20834. J Magn Reson Imaging. 2007. PMID: 17279535
-
Calculation of T2 relaxation time from ultrafast single shot sequences for differentiation of liver tumors: comparison of echo-planar, HASTE, and spin-echo sequences.Radiat Med. 2000 Jan-Feb;18(1):7-14. Radiat Med. 2000. PMID: 10852650
-
Diffusion-weighted MR of the brain: methodology and clinical application.Radiol Med. 2005 Mar;109(3):155-97. Radiol Med. 2005. PMID: 15775887 Review. English, Italian.
-
1H MR spectroscopy of the brain: absolute quantification of metabolites.Radiology. 2006 Aug;240(2):318-32. doi: 10.1148/radiol.2402050314. Radiology. 2006. PMID: 16864664 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources