Implantable hearing device performance measured by laser Doppler interferometry
- PMID: 9170711
Implantable hearing device performance measured by laser Doppler interferometry
Abstract
Recent application of the Doppler principle laser interferometry to audiology, acoustics and otology has facilitated the development of implantable hearing devices (IHDs). During the design and testing of two different electromagnetic middle ear implants for sensorineural hearing loss, we used single-point laser Doppler interferometry (LDI). A commercially available interferometer, internally calibrated and validated against a National Institute for Standards and Technology (NIST) standard, was used with both mechanical fixtures and fresh temporal bones to evaluate implant mass, shape and orientation, attachment, electromagnetic coupling and acoustic properties. At both Hough Ear Institute and Symphonix Devices, Inc., we have shown that high fidelity and amplitudes can be recorded in vitro over a frequency range of 500 Hz to 10 kHz. These data can provide greater assurance of safety and efficacy to regulatory agencies before entering clinical trials. We propose that LDI be considered as an international standard for accurate, consistent comparison of performances of all IHDs during development. Furthermore, the future availability of human IHD data will allow for the extrapolation of a mechanical bench model of the middle ear transfer function for use in quality control during manufacturing and diagnosis of failure in IHDs.
Similar articles
-
Mechanical, acoustic and electromagnetic evaluation of the semi-implantable middle ear hearing device (SIMEHD).Ear Nose Throat J. 1997 May;76(5):321-7. Ear Nose Throat J. 1997. PMID: 9170714
-
The middle ear bioelectronic microphone for a totally implantable cochlear hearing device for profound and total hearing loss.Am J Otol. 1999 Sep;20(5):602-11. Am J Otol. 1999. PMID: 10503582
-
In vitro model for intraoperative adjustments in an implantable hearing aid (MET).Laryngoscope. 2006 Mar;116(3):473-81. doi: 10.1097/01.MLG.0000199889.61483.8F. Laryngoscope. 2006. PMID: 16540912
-
Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.Laryngoscope. 2006 Aug;116(8):1321-33. doi: 10.1097/01.mlg.0000230479.39551.4a. Laryngoscope. 2006. PMID: 16885731 Review.
-
Implantable middle ear hearing devices: current state of technology and market challenges.Curr Opin Otolaryngol Head Neck Surg. 2006 Oct;14(5):314-8. doi: 10.1097/01.moo.0000244187.66807.30. Curr Opin Otolaryngol Head Neck Surg. 2006. PMID: 16974143 Review.
Cited by
-
Recent advances: otolaryngology.BMJ. 1999 Oct 23;319(7217):1110-3. doi: 10.1136/bmj.319.7217.1110. BMJ. 1999. PMID: 10531105 Free PMC article. Review. No abstract available.
-
The development of active middle ear implants: A historical perspective and clinical outcomes.Laryngoscope Investig Otolaryngol. 2018 Oct 6;3(5):394-404. doi: 10.1002/lio2.215. eCollection 2018 Oct. Laryngoscope Investig Otolaryngol. 2018. PMID: 30410994 Free PMC article.
-
[Semi-implantable hearing aids for sensorineural hearing loss and combined hearing loss: experiences at the German Armed Forces Hospital in Ulm].HNO. 2009 Mar;57(3):208-15. doi: 10.1007/s00106-008-1851-6. HNO. 2009. PMID: 19247625 Review. German.
-
A longitudinal study on postoperative hearing thresholds with the Vibrant Soundbridge device.Eur Arch Otorhinolaryngol. 2004 Oct;261(9):493-6. doi: 10.1007/s00405-003-0669-9. Eur Arch Otorhinolaryngol. 2004. PMID: 15004703
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical