Comparison of pressure- and flow-triggered pressure-support ventilation on weaning parameters in patients recovering from acute respiratory failure
- PMID: 9187592
- DOI: 10.1097/00003246-199705000-00008
Comparison of pressure- and flow-triggered pressure-support ventilation on weaning parameters in patients recovering from acute respiratory failure
Abstract
Objective: To compare the effects of pressure- and flow-triggered pressure-support ventilation on weaning parameters during recovery from acute respiratory failure.
Design: Prospective, randomized, clinical trial.
Setting: Intensive care unit in a university hospital.
Patients: Sixteen orotracheally intubated adult patients recovering from acute respiratory failure of various etiologies, without chronic obstructive pulmonary disease.
Interventions: Randomized application of pressure- and flow-triggered pressure-support ventilation at 100% and 75% ventilatory support levels in each triggering system. A total of four conditions were applied for 30 mins each in all patients.
Measurements and main results: Ventilatory, respiratory, and hemodynamic data were measured. For the measurement of weaning parameters, pressure and volume signals were directed to a computerized respiratory monitor by means of an esophageal probe and a flow sensor between the "Y" piece of the ventilatory circuit and the endotracheal tube. During both pressure-triggered (trigger sensitivity of -1 cm H2O) and flow-triggered (trigger sensitivity of 0.7 to 2.0 L/min) pressure-support ventilation with a ventilator, peak airway pressures were applied so as to decrease the work of breathing performed by the patient to zero (full ventilatory support). Partial ventilatory support was applied at 75% of the peak airway pressures achieved during full ventilatory support with each triggering system. A total of four experimental conditions were evaluated at identical FiO2 and positive and-expiratory pressure levels during pressure-support ventilation in each patient. Total ventilation volumes, arterial blood gas data, and hemodynamics did not differ among the four experimental conditions. During partial ventilatory support, the work of breathing, rapid shallow breathing index, and esophageal pressure increased significantly with both triggering systems when compared with data obtained at full ventilatory support. The mean data for the weaning parameters during the condition of partial ventilatory support were comparable between pressure- and flow-triggered pressure-support ventilation (i.e., 0.38 +/- 0.24 vs. 0.42 +/- 0.26 joule/L for work of breathing, 2.6 +/- 1.6 vs. 3.3 +/- 1.7 cm H2O for tracheal occlusion pressure, and 40.2 +/- 12.9 vs. 50.4 +/- 18.3 breaths/min/L for rapid shallow breathing index, respectively).
Conclusions: The application of either a pressure- or flow-triggered system during pressure-support ventilation with the ventilator did not significantly affect short-term changes in gas exchange, respiratory mechanics, and inspiratory workload in patients recovering from acute respiratory failure of various etiologies without chronic obstructive pulmonary disease.
Similar articles
-
Comparisons of predictive performance of breathing pattern variability measured during T-piece, automatic tube compensation, and pressure support ventilation for weaning intensive care unit patients from mechanical ventilation.Crit Care Med. 2011 Oct;39(10):2253-62. doi: 10.1097/CCM.0b013e31822279ed. Crit Care Med. 2011. PMID: 21666447
-
Respiratory muscle performance, pulmonary mechanics, and gas exchange between the BiPAP S/T-D system and the Servo Ventilator 900C with bilevel positive airway pressure ventilation following gradual pressure support weaning.Chest. 1998 Nov;114(5):1390-6. doi: 10.1378/chest.114.5.1390. Chest. 1998. PMID: 9824020 Clinical Trial.
-
Intratracheal pulmonary ventilation in a rabbit lung injury model: continuous airway pressure monitoring and gas exchange efficacy.Crit Care Med. 2000 Jul;28(7):2480-5. doi: 10.1097/00003246-200007000-00049. Crit Care Med. 2000. PMID: 10921582
-
Ventilatory failure, ventilator support, and ventilator weaning.Compr Physiol. 2012 Oct;2(4):2871-921. doi: 10.1002/cphy.c110030. Compr Physiol. 2012. PMID: 23720268 Review.
-
The impact of spontaneous breathing during mechanical ventilation.Curr Opin Crit Care. 2006 Feb;12(1):13-8. doi: 10.1097/01.ccx.0000198994.37319.60. Curr Opin Crit Care. 2006. PMID: 16394778 Review.
Cited by
-
Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling - a case series and literature review.J Cardiothorac Surg. 2010 May 17;5:39. doi: 10.1186/1749-8090-5-39. J Cardiothorac Surg. 2010. PMID: 20478064 Free PMC article.
-
Weaning from mechanical ventilation.Crit Care. 2000;4(2):72-80. doi: 10.1186/cc660. Epub 2000 Feb 18. Crit Care. 2000. PMID: 11094496 Free PMC article. Review.
-
Clinical review: patient-ventilator interaction in chronic obstructive pulmonary disease.Crit Care. 2006;10(6):236. doi: 10.1186/cc5073. Crit Care. 2006. PMID: 17096868 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources