Comparative Study
. 1997 Jul-Aug;21(4):554-66.
doi: 10.1097/00004728-199707000-00007.

Comparison and evaluation of retrospective intermodality brain image registration techniques

Affiliations
Free article
Comparative Study

Comparison and evaluation of retrospective intermodality brain image registration techniques

J West et al. J Comput Assist Tomogr. 1997 Jul-Aug.
Free article

Abstract

Purpose: The primary objective of this study is to perform a blinded evaluation of a group of retrospective image registration techniques using as a gold standard a prospective, marker-based registration method. To ensure blindedness, all retrospective registrations were performed by participants who had no knowledge of the gold standard results until after their results had been submitted. A secondary goal of the project is to evaluate the importance of correcting geometrical distortion in MR images by comparing the retrospective registration error in the rectified images, i.e., those that have had the distortion correction applied, with that of the same images before rectification.

Method: Image volumes of three modalities (CT, MR, and PET) were obtained from patients undergoing neurosurgery at Vanderbilt University Medical Center on whom bone-implanted fiducial markers were mounted. These volumes had all traces of the markers removed and were provided via the Internet to project collaborators outside Vanderbilt, who then performed retrospective registrations on the volumes, calculating transformations from CT to MR and/ or from PET to MR. These investigators communicated their transformations again via the Internet to Vanderbilt, where the accuracy of each registration was evaluated. In this evaluation, the accuracy is measured at multiple volumes of interest (VOIs), i.e., areas in the brain that would commonly be areas of neurological interest. A VOI is defined in the MR image and its centroid c is determined. Then, the prospective registration is used to obtain the corresponding point c' in CT or PET. To this point, the retrospective registration is then applied, producing c" in MR. Statistics are gathered on the target registration error (TRE), which is the distance between the original point c and its corresponding point c".

Results: This article presents statistics on the TRE calculated for each registration technique in this study and provides a brief description of each technique and an estimate of both preparation and execution time needed to perform the registration.

Conclusion: Our results indicate that retrospective techniques have the potential to produce satisfactory results much of the time, but that visual inspection is necessary to guard against large errors.