Distortion of intracardiac electrograms following defibrillator shocks for atrial tachyarrhythmias
- PMID: 9227767
- DOI: 10.1111/j.1540-8159.1997.tb03539.x
Distortion of intracardiac electrograms following defibrillator shocks for atrial tachyarrhythmias
Abstract
In three patients with a defibrillator system consisting of a Ventak P2 pulse generator and an Endotak C transvenous lead, we observed distortion of intracardiac electrograms following defibrillator shocks for atrial arrhythmias. There was a transient marked widening of the intracardiac ventricular complexes resembling ventricular tachycardia. This phenomenon should be recognized when evaluating arrhythmic episodes.