Long-term in vivo alterations of polyester vascular grafts in humans
- PMID: 9236706
- DOI: 10.1016/s1078-5884(97)80062-7
Long-term in vivo alterations of polyester vascular grafts in humans
Abstract
Objectives: To examine the influence of in vivo hydrolysis on the physical properties of polyester grafts and their correlation to the period of implantation in the human body.
Materials and methods: Sixty-five explanted vascular grafts were obtained after 0-23 years of implantation due to suture aneurysms (18), occlusion (12), graft infection (12), failure of graft material (7) and post-mortem (16). The surface was examined by scanning electron microscopy, the molecular integrity by infra-red spectroscopy and physical strength by probe puncture.
Results: Scission of macromolecular chains and loss of strength were shown. It was demonstrated that hydrolytic degradation of polyester takes place with increasing time of implantation in humans. Analysis by linear regression showed that polyester grafts lose 31.4% of their bursting strength in 10 years and 100% in 25-39 years after implantation.
Conclusions: Regular follow-ups of patients with aged vascular grafts and the precise documentation of implanted materials are necessary to estimate graft degradation.
Similar articles
-
Long-term dilatation of polyester and expanded polytetrafluoroethylene tube grafts after open repair of infrarenal abdominal aortic aneurysms.J Vasc Surg. 2011 Jun;53(6):1506-13. doi: 10.1016/j.jvs.2011.02.028. Epub 2011 May 4. J Vasc Surg. 2011. PMID: 21536404 Clinical Trial.
-
The effect of autoclave resterilisation on polyester vascular grafts.Eur J Vasc Endovasc Surg. 1999 Nov;18(5):386-90. doi: 10.1053/ejvs.1999.0891. Eur J Vasc Endovasc Surg. 1999. PMID: 10610826
-
In vitro and in vivo characterization of a silk fibroin-coated polyester vascular prosthesis.Artif Organs. 2008 Dec;32(12):932-41. doi: 10.1111/j.1525-1594.2008.00655.x. Artif Organs. 2008. PMID: 19133021
-
[The Dacron vascular prosthesis in men. Mechanical performance, lesions and molecular stability].Presse Med. 1990 Apr 14;19(15):709-14. Presse Med. 1990. PMID: 2139962 Review. French.
-
Pathology of graft and stent-graft infections: Lessons learned from examination of explant materials.Semin Vasc Surg. 2017 Jun-Sep;30(2-3):70-74. doi: 10.1053/j.semvascsurg.2017.10.002. Epub 2017 Oct 22. Semin Vasc Surg. 2017. PMID: 29248122 Review.
Cited by
-
Collagen foil (TissuFoil E) reduces the formation of adhesions when using polypropylene mesh for the repair of experimental abdominal wall defects.Int J Colorectal Dis. 2006 Dec;21(8):840-6. doi: 10.1007/s00384-006-0091-z. Epub 2006 Mar 7. Int J Colorectal Dis. 2006. PMID: 16520931
-
An evaluation of the host response to an interspinous process device based on a series of spine explants: Device for Intervertebral Assisted Motion (DIAM®).J Spine Surg. 2019 Dec;5(4):483-495. doi: 10.21037/jss.2019.10.13. J Spine Surg. 2019. PMID: 32042999 Free PMC article.
-
Improving Biocompatibility of Polyester Fabrics through Polyurethane/Gelatin Complex Coating for Potential Vascular Application.Polymers (Basel). 2022 Feb 28;14(5):989. doi: 10.3390/polym14050989. Polymers (Basel). 2022. PMID: 35267812 Free PMC article.
-
Materials characterization and histological analysis of explanted polypropylene, PTFE, and PET hernia meshes from an individual patient.J Mater Sci Mater Med. 2013 Apr;24(4):1113-22. doi: 10.1007/s10856-013-4872-y. Epub 2013 Jan 31. J Mater Sci Mater Med. 2013. PMID: 23371769 Free PMC article.
-
Biocompatibility of prosthetic meshes in abdominal surgery.Semin Immunopathol. 2011 May;33(3):235-43. doi: 10.1007/s00281-010-0237-5. Epub 2011 Jan 12. Semin Immunopathol. 2011. PMID: 21225262 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources