Incision sizes for foldable intraocular lenses
- PMID: 9261314
- DOI: 10.1016/s0161-6420(97)30147-x
Incision sizes for foldable intraocular lenses
Abstract
Objective: The purpose of the study is to determine the minimal incision sizes required for implantation of a variety of different foldable intraocular lenses (IOLs) and to evaluate the effect of incision size on tissue damage.
Design: Randomized experimental study.
Participants: Sixty-nine fresh human cadaveric eyes: 15 (pilot study), 48 (main study), and 6 (scanning electron microscopy study).
Intervention: Implantation of foldable IOLs into cadaveric eyes.
Main outcome measures: In 48 fresh human cadaveric eyes, limbal corneal tunnel incisions were made, and in a randomized fashion, 8 different foldable IOLs of 20.5 diopters (D) were inserted: 4 silicone (SI-30NB, Allergan Medical Optics, Irvine, CA; C10UB, Chiron Ophthalmics, Inc, Irvine, CA; LI41U, IOLAB, Chiron Ophthalmics, Inc, Irvine, CA; AA-4203, Staar Surgical Company, Monrovia, CA); two soft acrylic (MA60BM and MA30BA, both Alcon Laboratories, Inc, Ft. Worth, TX); and two hydrogel models (SH30BC, Alcon; H60M, Storz Ophthalmics, Inc, St. Louis, MO). For each IOL model, six insertions were performed with a recommended implantation device. Using calipers, the authors measured internal and external incision sizes before and after insertion. Scanning electron microscopy was performed on selected incisions in six additional human cadaveric eyes.
Results: Incision sizes after insertion ranged from 3.2 to 3.8 mm. The smallest incisions permitting IOL insertion were associated with the injectors. However, these incisions enlarged after insertion by approximately 11% and then were similar to the incision sizes after forceps insertion of the high-refractive-index silicone, the 5.5-mm optic acrylic, and the one-piece hydrogel IOL. The largest incisions were associated with the 6-mm acrylic IOL and the three-piece silicone IOL with a lower refractive index. The scanning electron microscopy showed tearing of corneal tissue after implantation through the smallest incision; this was more pronounced with injectors than with forceps.
Conclusions: Corneal tunnel incisions enlarged up to 11% after insertion of foldable IOLs through the smallest possible incision. With current technology, the smallest postinsertion incision size of a 20.5-D foldable IOL is 3.2 mm.
Similar articles
-
Experimental and clinical evaluation of incision size and shape following forceps and injector implantation of a three-piece high-refractive-index silicone intraocular lens.Graefes Arch Clin Exp Ophthalmol. 1998 Dec;236(12):922-8. doi: 10.1007/s004170050181. Graefes Arch Clin Exp Ophthalmol. 1998. PMID: 9865623 Clinical Trial.
-
Incision sizes with 5.5 mm total optic, 3-piece foldable intraocular lenses.J Cataract Refract Surg. 2000 Dec;26(12):1765-72. doi: 10.1016/s0886-3350(00)00526-5. J Cataract Refract Surg. 2000. PMID: 11134877 Clinical Trial.
-
Incision width after phacoemulsification with foldable intraocular lens implantation.J Cataract Refract Surg. 2000 Feb;26(2):237-41. doi: 10.1016/s0886-3350(99)00346-6. J Cataract Refract Surg. 2000. PMID: 10683791
-
Incision sizes before and after implantation of 6-mm optic foldable intraocular lenses using Monarch and Unfolder injector systems.Ophthalmology. 2005 Jan;112(1):58-66. doi: 10.1016/j.ophtha.2004.06.030. Ophthalmology. 2005. PMID: 15629821 Review.
-
[Sutureless scleral intraocular lens fixation: report of nine cases and literature review].J Fr Ophtalmol. 2013 Oct;36(8):658-68. doi: 10.1016/j.jfo.2012.09.009. Epub 2013 Jul 25. J Fr Ophtalmol. 2013. PMID: 23891322 Review. French.
Cited by
-
Clear corneal incision, an important step in modern cataract surgery: a review.Eye (Lond). 2023 Oct;37(14):2864-2876. doi: 10.1038/s41433-023-02440-z. Epub 2023 Feb 14. Eye (Lond). 2023. PMID: 36788364 Free PMC article. Review.
-
Corneal incision architecture after IOL implantation with three different injectors: an environmental scanning electron microscopy study.Int Ophthalmol. 2019 Feb;39(2):397-403. doi: 10.1007/s10792-018-0825-2. Epub 2018 Feb 1. Int Ophthalmol. 2019. PMID: 29392640
-
Wound Integrity of 2.0 mm Transconjunctival Single-Plane Sclerocorneal Incision: A Comparison between a Motorized Injector with/without Pause Time and a Manual Injector.J Ophthalmol. 2016;2016:8129298. doi: 10.1155/2016/8129298. Epub 2016 Feb 17. J Ophthalmol. 2016. PMID: 26989502 Free PMC article.
-
Comparison of surgically induced astigmatisms after clear corneal incisions of different sizes.Korean J Ophthalmol. 2007 Mar;21(1):1-5. doi: 10.3341/kjo.2007.21.1.1. Korean J Ophthalmol. 2007. PMID: 17460424 Free PMC article.
-
Scanning electron microscopy study of different one-piece foldable acrylic intraocular lenses after injection through microincisional cataract surgery cartridges.Int Ophthalmol. 2020 Feb;40(2):369-376. doi: 10.1007/s10792-019-01193-7. Epub 2019 Oct 17. Int Ophthalmol. 2020. PMID: 31624987
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources