Comparative Study
. 1997 Sep;121(9):956-9.

Effect of citrate concentration in specimen collection tubes on the International Normalized Ratio

Affiliations
  • PMID: 9302927
Comparative Study

Effect of citrate concentration in specimen collection tubes on the International Normalized Ratio

C F Danielson et al. Arch Pathol Lab Med. 1997 Sep.

Erratum in

  • Arch pathol Lab Med 1997 Dec;121(12):2502

Abstract

Problem: Does citrate concentration in specimen collection tubes affect the International Normalized Ratio?

Methods: The International Normalized Ratio was determined on quadruplicate plasma specimens from 32 patients treated long term with oral anticoagulants-two from tubes with 3.2% citrate and two with 3.8% citrate. Two laboratories, using two different coagulometers, tested the specimens.

Results: International Normalized Ratios of plasma from tubes with 3.8% citrate were significantly higher than those from tubes with 3.2% citrate when tested with either coagulometer. Patients given adequate anticoagulation on the basis of the International Normalized Ratio at one concentration of citrate appeared either overanticoagulated and at risk of bleeding or underanticoagulated and at risk of thromboembolism at the other concentration of citrate.

Conclusion: Results emphasize the need for using a single concentration of citrate for prothrombin time testing. We recommend 3.2% citrate.