Inaccuracy of portable peak flow meters: correction is not needed. Dutch CNSLD Study Group
- PMID: 9307172
- DOI: 10.1111/j.1651-2227.1997.tb08616.x
Inaccuracy of portable peak flow meters: correction is not needed. Dutch CNSLD Study Group
Abstract
This study examined whether correction of peak expiratory flow (PEF) values for the inaccuracy of the meter would affect asthma management in 102 children (7-14 y old). PEF was recorded with a mini Wright meter twice daily for 2 weeks. As expected, measured PEF overestimated PEF level and asthma control in these children on many diary days. The actual numerical differences between measured and corrected PEF on these days were very small (>5% in only five patients, maximum 10%). It is unlikely that such small changes in PEF justify changes in asthma management, even if these changes cause PEF levels to cross arbitrary borders between various levels of asthma control used in self-management plans. The clinical importance of the inaccuracy of portable PEF meters is negligible.
Similar articles
-
Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan.Thorax. 1996 Apr;51(4):403-6. doi: 10.1136/thx.51.4.403. Thorax. 1996. PMID: 8733493 Free PMC article.
-
Changes in peak expiratory flow indices as a proxy for changes in bronchial hyperresponsiveness. Dutch Chronic Non-Specific Lung Disease study group.Eur Respir J. 2000 Aug;16(2):220-5. doi: 10.1034/j.1399-3003.2000.16b07.x. Eur Respir J. 2000. PMID: 10968495 Clinical Trial.
-
[Impact of inaccuracy of peak flowmeters with Wright scale on asthma severity assessment in children].Med Pregl. 2008 Jul-Aug;61(7-8):354-8. doi: 10.2298/mpns0808354s. Med Pregl. 2008. PMID: 19097371 Serbian.
-
Incorrect use of peak flow meters: are you observing your patients?J Asthma. 2014 Aug;51(6):566-72. doi: 10.3109/02770903.2014.914218. Epub 2014 May 9. J Asthma. 2014. PMID: 24720711 Review.
-
Usefulness of monitoring lung function in asthma.Arch Dis Child. 2003 Nov;88(11):1021-5. doi: 10.1136/adc.88.11.1021. Arch Dis Child. 2003. PMID: 14612375 Free PMC article. Review.
Cited by
-
eHealth Technologies for Monitoring Pediatric Asthma at Home: Scoping Review.J Med Internet Res. 2023 Jul 21;25:e45896. doi: 10.2196/45896. J Med Internet Res. 2023. PMID: 37477966 Free PMC article.
-
Peak flow variation in childhood asthma: correlation with symptoms, airways obstruction, and hyperresponsiveness during long-term treatment with inhaled corticosteroids. Dutch CNSLD Study Group.Thorax. 1999 Feb;54(2):103-7. doi: 10.1136/thx.54.2.103. Thorax. 1999. PMID: 10325912 Free PMC article. Clinical Trial.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials