Analysis of acoustic noise in MRI
- PMID: 9309612
- DOI: 10.1016/s0730-725x(97)00090-8
Analysis of acoustic noise in MRI
Abstract
Acoustic or sound noise due to gradient pulsing has been one of the problems in magnetic resonance imaging (MRI), both in patient scanning as well as in many areas of psychiatric and neuroscience research such as functional MRI. Our recent observations in functional MRI for the visual and motor cortex show very different results with sound noise in comparison with the results obtained without sound noise. Although a number of ideas have been suggested in the literature about the possible elimination or reduction of sound noise, progress has been slow due to the basic role of gradient pulsing in MR imaging. Before we tackle the sound-noise-reduction problem, we believe that a systematic study of sound or acoustic noise behavior will provide important information for future endeavors in this area of research in MRI systems, in both commercial and research systems. Therefore, we report on some typical behavior of sound noise observed from MRI scanners and the analyses of their characteristics. Data are obtained both from a commercial MRI scanner (GE Signa 1.5-T EPI system) as well as a research-type MRI scanner (KAIS 2.0-T) developed at a university laboratory setting.
Similar articles
-
Effects of the acoustic noise of the gradient systems on fMRI: a study on auditory, motor, and visual cortices.Magn Reson Med. 1998 Feb;39(2):331-5. doi: 10.1002/mrm.1910390224. Magn Reson Med. 1998. PMID: 9469720
-
Quiet MRI with novel acoustic noise reduction.MAGMA. 2002 Jan;13(3):139-44. doi: 10.1007/BF02678588. MAGMA. 2002. PMID: 11755088
-
Analysis of magnetic resonance imaging acoustic noise generated by a 4.7 T experimental system.Acta Otolaryngol. 2000 Sep;120(6):739-43. doi: 10.1080/000164800750000270. Acta Otolaryngol. 2000. PMID: 11099151
-
Advances in Acoustic Signal Processing Techniques for Enhanced Bowel Sound Analysis.IEEE Rev Biomed Eng. 2019;12:240-253. doi: 10.1109/RBME.2018.2874037. Epub 2018 Oct 11. IEEE Rev Biomed Eng. 2019. PMID: 30307875 Review.
-
Acoustic noise and functional magnetic resonance imaging: current strategies and future prospects.J Magn Reson Imaging. 2002 Nov;16(5):497-510. doi: 10.1002/jmri.10186. J Magn Reson Imaging. 2002. PMID: 12412026 Review.
Cited by
-
Extraction of overt verbal response from the acoustic noise in a functional magnetic resonance imaging scan by use of segmented active noise cancellation.Magn Reson Med. 2005 Mar;53(3):739-44. doi: 10.1002/mrm.20398. Magn Reson Med. 2005. PMID: 15723385 Free PMC article.
-
Making MR imaging child's play - pediatric neuroimaging protocol, guidelines and procedure.J Vis Exp. 2009 Jul 30;(29):1309. doi: 10.3791/1309. J Vis Exp. 2009. PMID: 19684560 Free PMC article.
-
Individualised MRI training for paediatric neuroimaging: A child-focused approach.Dev Cogn Neurosci. 2020 Feb;41:100750. doi: 10.1016/j.dcn.2019.100750. Epub 2019 Dec 16. Dev Cogn Neurosci. 2020. PMID: 31999567 Free PMC article.
-
Overview of Methods for Noise and Heat Reduction in MRI Gradient Coils.Front Phys. 2022 Jul;10:907619. doi: 10.3389/fphy.2022.907619. Epub 2022 Jul 8. Front Phys. 2022. PMID: 36506821 Free PMC article.
-
fMRI Mapping of Brain Activity Associated with the Vocal Production of Consonant and Dissonant Intervals.J Vis Exp. 2017 May 23;(123):55419. doi: 10.3791/55419. J Vis Exp. 2017. PMID: 28570522 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical