Radiation exposure to patients undergoing diagnostic and interventional cardiac catheterization procedures
- PMID: 9328690
- DOI: 10.1002/(sici)1097-0304(199710)42:2<121::aid-ccd4>3.0.co;2-f
Radiation exposure to patients undergoing diagnostic and interventional cardiac catheterization procedures
Abstract
There has been a recent wave of concern over radiation exposure to patients during cardiac catheterization. Accordingly, we measured the area-exposure product (AEP) to patients undergoing diagnostic and interventional cardiac catheterization procedures. Fluoroscopic and cinefluorographic exposures were determined for 510 patients using an AEP meter. The total AEP was higher in interventional than diagnostic cases (16,289 R.cm2 vs. 10,873 R.cm2, P < 0.00001); multi-lesion than single lesion interventions (20,311 R.cm2 vs. 15,919 R.cm2 P < 0.0001); and in patients with previous coronary bypass surgery (20,403 R.cm2 vs. 14,298 R.cm2, P < 0.00001). The highest AEPs were observed in patients with a prior history of bypass surgery who underwent diagnostic catheterization and multilesion intervention during the same procedure.
Comment in
-
Radiation exposure: clueless in the cath lab, or sayonara ALARA.Cathet Cardiovasc Diagn. 1997 Oct;42(2):126-7. doi: 10.1002/(sici)1097-0304(199710)42:2<126::aid-ccd5>3.0.co;2-f. Cathet Cardiovasc Diagn. 1997. PMID: 9328691 No abstract available.
-
X-ray-ted.Cathet Cardiovasc Diagn. 1997 Oct;42(2):128-9. doi: 10.1002/(sici)1097-0304(199710)42:2<128::aid-ccd6>3.0.co;2-f. Cathet Cardiovasc Diagn. 1997. PMID: 9328692 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources