Coil on the loop and selectively enhanced stiffness for improved control: feasibility study in the neonatal ovine model of the large patent ductus arteriosus
- PMID: 9342124
- DOI: 10.1097/00004424-199710000-00009
Coil on the loop and selectively enhanced stiffness for improved control: feasibility study in the neonatal ovine model of the large patent ductus arteriosus
Abstract
Rationale and objectives: Several devices have been suggested and tested for interventional closure of the persistent ductus arteriosus. Coils were found effective only in small ducts due to their lack of maneuverability and physical limitation of grip forces leading to risk of embolization. The authors evaluated the feasibility to place single coils with selectively enhanced stiffness into high shunting ductus arteriosus, the coils being deployed and controlled through a veno-arterial loop in a bovine model.
Methods: "Double-cone" shaped, stainless steel coils with enhanced stiffness of the outer rings were mounted on either end on a nitinol core wire. A snap-in mechanism attaches the coil to this delivery wire and is freed by a pusher system of coiled steel wire that is advanced over the core wire. Forming a veno-arterial loop across the patent ductus allows for repositioning into the pulmonary artery or aortic catheter. A chronic lamb model of large patent ductus arteriosus (PDAs) (> or = 9 mm) was used in which ductus patency was secured by a protocol of repetitive angioplasties. Different systems (n = 10) were placed having retrieved the previous coil by a snare after definitive release.
Results: Placement of coils was possible in all 10 attempts. The coils were pulled back into the catheters between five and eight times before definitive release.
Conclusions: The device allows controlled placement of single coils in our model of large PDAs and may lead the way to overcome previous limitations of coils. Clinical trials are warranted.
Similar articles
-
Double-helix coil for occlusion of large patent ductus arteriosus: evaluation in a chronic lamb model.J Am Coll Cardiol. 1998 Mar 1;31(3):677-83. doi: 10.1016/s0735-1097(98)00025-4. J Am Coll Cardiol. 1998. PMID: 9502653
-
A small interventional device to occlude persistently patent ductus arteriosus in neonates: evaluation in piglets.J Am Coll Cardiol. 1996 Oct;28(4):1024-30. doi: 10.1016/s0735-1097(96)00242-2. J Am Coll Cardiol. 1996. PMID: 8837585
-
Occlusion of the neonatal patent ductus arteriosus with a simple retrievable device: a feasibility study.Cardiovasc Intervent Radiol. 1996 May-Jun;19(3):170-5. doi: 10.1007/BF02577614. Cardiovasc Intervent Radiol. 1996. PMID: 8661644
-
[Transcatheter treatment of patent ductus arteriosus].Ital Heart J Suppl. 2002 Nov;3(11):1092-7. Ital Heart J Suppl. 2002. PMID: 12506510 Review. Italian.
-
Transarterial coil embolization for canine patent ductus arteriosus occlusion.Clin Tech Small Anim Pract. 2005 Aug;20(3):196-202. doi: 10.1053/j.ctsap.2005.05.008. Clin Tech Small Anim Pract. 2005. PMID: 16180402 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources