Comparative Study
. 1997 Sep;23(9):987-91.
doi: 10.1007/s001340050443.

Clinical validation of cardiac output measurements using femoral artery thermodilution with direct Fick in ventilated children and infants

Affiliations
Comparative Study

Clinical validation of cardiac output measurements using femoral artery thermodilution with direct Fick in ventilated children and infants

S M Tibby et al. Intensive Care Med. 1997 Sep.

Abstract

Objective: To validate clinically cardiac output (CO) measurements using femoral artery thermodilution in ventilated children and infants by comparison with CO estimated from the Fick equation via a metabolic monitor.

Design: Prospective, comparison study.

Setting: Paediatric intensive care unit of a university hospital.

Patients: 24 ventilated infants and children, aged 0.3 to 175 months (median age 19 months).

Interventions: Oxygen consumption measurements were made and averaged over a 5-min period, at the end of which arterial and mixed venous blood samples were taken and oxygen saturations measured by co-oximetry, with CO being calculated using the Fick equation. Over this 5-min period, five sets of femoral arterial thermodilution (FATD) measurements were made and averaged. One comparison of CO values was made per patient.

Results: Mean Fick CO was 2.55 l/min (range 0.24 to 8.71 l/min) and mean FATD CO was 2.51 l/min (range 0.28-7.96 l/min). The mean bias was 0.03 l/min (95% confidence interval -0.07 to 0.14 l/min), with limits of agreement of -0.45 to 0.52 l/min. When indexed to body surface area, the mean Fick cardiac index became 3.51 l/min per m2 (1.52-6.98 l/min per m2) and mean FATD 3.49 l/min per m2 (1.74-6.84 l/min per m2). The mean bias was 0.02 l/min per m2 (95% confidence interval -0.11 to 0.15 l/min per m2) with limits of agreement of-0.57 to 0.61 l/min per m2. The mean FATD coefficient of variation was 5.8% (SEM 0.5%).

Conclusions: FATD compares favourably with Fick derived CO estimates in infants and children and may represent an advance in haemodynamic monitoring of critically ill children.