Comparison of serum protein electrophoresis by agarose gel and capillary zone electrophoresis in a clinical setting
- PMID: 9372270
- DOI: 10.1002/elps.1150181012
Comparison of serum protein electrophoresis by agarose gel and capillary zone electrophoresis in a clinical setting
Abstract
Capillary zone electrophoresis (CZE) offers the potential for automating serum protein electrophoretic analysis traditionally performed on standard thin-layer agarose gels. The following describes the use of CZE compared to agarose gel electrophoresis (AGE) for the detection of dysproteinemia and paraproteinemia in a clinical study involving 240 patients. The study includes within-run and between-run reproducibility data on the Paragon CZE 2000 Clinical Capillary Electrophoresis System, in addition to concordance data between the two methodologies. Paraprotein quantitation studies comparing AGE versus CZE were also performed. Reproducibility for the automated CZE system was superior to the AGE system. Improved reproducibility for the CZE method is largely due to measuring protein absorbance directly at 214 nm versus the traditional AGE method that measures the amount of dye adsorbed to protein. Reproducibility data as percent coefficient of variance (% CV) for the five classic bands in a normal control serum for between-run precision ranged from 1.2 to 4.5% for CZE compared to AGE, which ranged from 3.8 to 8.0% CV. Concordance studies between AGE and CZE involving dysproteinemias including hypogammaglobulinemia, polyclonal and monoclonal gammopathies, acute and chronic inflammation, nephrosis, hepatodegenerative disease, cirrhosis, and iron deficiency anemia showed 96% agreement. Paraprotein classification, which compared the CZE immunosubtraction method to immunofixation electrophoresis (IFE) on agarose, showed 100% agreement. Certain dysproteinemias involving beta lipoprotein were in partial concordance due to the inability of the CZE procedure to detect this component. Detection limits for monoclonal gammopathies, providing they were not comigrating with other proteins, were IgG 50 mg/dL, IgM 75 mg/dL, and IgA 75 mg/dL. Paraprotein quantitative studies between the two methods showed less than a +/- 0.2 g/dL variation.
Similar articles
-
Evaluation of a capillary zone electrophoresis system versus a conventional agarose gel system for routine serum protein separation and monoclonal component typing.Clin Lab. 2006;52(1-2):19-27. Clin Lab. 2006. PMID: 16506360
-
Performance of the Sebia CAPILLARYS 2 for detection and immunotyping of serum monoclonal paraproteins.Am J Clin Pathol. 2007 Aug;128(2):293-9. doi: 10.1309/1L3CG8GK6F8VYNYH. Am J Clin Pathol. 2007. PMID: 17638665
-
Evaluation of a new capillary zone electrophoresis system for the identification and typing of Bence Jones Protein.Clin Biochem. 2006 Feb;39(2):152-9. doi: 10.1016/j.clinbiochem.2005.10.016. Epub 2005 Dec 9. Clin Biochem. 2006. PMID: 16337932
-
Separation of serum proteins by automated capillary zone electrophoresis.Clin Chem Lab Med. 2003 Jun;41(6):762-72. doi: 10.1515/CCLM.2003.116. Clin Chem Lab Med. 2003. PMID: 12880139 Review.
-
Advances in serum protein electrophoresis.Adv Clin Chem. 2006;42:43-80. doi: 10.1016/s0065-2423(06)42002-3. Adv Clin Chem. 2006. PMID: 17131624 Review.
Cited by
-
Establishment of Reference Intervals for Serum Protein Electrophoresis of Apparently Healthy Adults in Addis Ababa, Ethiopia.Int J Gen Med. 2022 Aug 23;15:6701-6711. doi: 10.2147/IJGM.S376450. eCollection 2022. Int J Gen Med. 2022. PMID: 36039308 Free PMC article.
-
A Comparison of gel (Hydragel 30) and capillary (Capillarys III Tera) electrophoresis for the characterization of human serum proteins.Pract Lab Med. 2021 Apr 29;25:e00233. doi: 10.1016/j.plabm.2021.e00233. eCollection 2021 May. Pract Lab Med. 2021. PMID: 34095418 Free PMC article.
-
Detection of kappa and lambda light chain monoclonal proteins in human serum: automated immunoassay versus immunofixation electrophoresis.Clin Vaccine Immunol. 2006 Feb;13(2):277-80. doi: 10.1128/CVI.13.2.277-280.2006. Clin Vaccine Immunol. 2006. PMID: 16467338 Free PMC article.
-
Dual detection high-speed capillary electrophoresis for simultaneous serum protein analysis and immunoassays.Sci Rep. 2022 Feb 4;12(1):1951. doi: 10.1038/s41598-022-05956-8. Sci Rep. 2022. PMID: 35121780 Free PMC article.
-
Quantification of peptides from immunoglobulin constant and variable regions by LC-MRM MS for assessment of multiple myeloma patients.Proteomics Clin Appl. 2014 Oct;8(9-10):783-95. doi: 10.1002/prca.201300077. Epub 2014 Sep 15. Proteomics Clin Appl. 2014. PMID: 24723328 Free PMC article.