Long-term quality control of DXA: a comparison of Shewhart rules and Cusum charts
- PMID: 9373567
- DOI: 10.1007/BF01623774
Long-term quality control of DXA: a comparison of Shewhart rules and Cusum charts
Abstract
Long-term clinical trials using bone mineral density (BMD) measurements are now commonplace. It is important to maintain a high standard of quality control. In this study two methods of monitoring quality control measurements of dual-energy X-ray absorptiometry (DXA) equipment have been compared using receiver operating curve (ROC) analysis: Shewhart multi-rule charts and Cusum charts. Computer-generated faults within daily spine phantom measurements were used. The Charts were then applied to 3 years of quality control data from one machine and the results related to hardware faults, random events and changes in the underlying BMD measurement. In the ROC analysis the Shewhart multi-rule chart performed as well or better than the cusum chart, but the stringency of the rules must be tightened to obtain optimal performance.
Similar articles
-
DXA longitudinal quality control: a comparison of inbuilt quality assurance, visual inspection, multi-rule Shewhart charts and Cusum analysis.Osteoporos Int. 1997;7(3):231-7. doi: 10.1007/BF01622294. Osteoporos Int. 1997. PMID: 9205636
-
Dual X-ray absorptiometry quality control: comparison of visual examination and process-control charts.J Bone Miner Res. 1996 May;11(5):626-37. doi: 10.1002/jbmr.5650110510. J Bone Miner Res. 1996. PMID: 9157777
-
Retrospective evaluation and adjustment of dual energy X-ray absorptiometry measurements for bone mineral density research studies.J Clin Densitom. 2002 Winter;5(4):421-33. doi: 10.1385/jcd:5:4:421. J Clin Densitom. 2002. PMID: 12665643
-
Comparison of control charts for monitoring clinical performance using binary data.BMJ Qual Saf. 2017 Nov;26(11):919-928. doi: 10.1136/bmjqs-2016-005526. Epub 2017 Sep 25. BMJ Qual Saf. 2017. PMID: 28947635 Free PMC article. Review.
-
Technical principles of dual energy x-ray absorptiometry.Semin Nucl Med. 1997 Jul;27(3):210-28. doi: 10.1016/s0001-2998(97)80025-6. Semin Nucl Med. 1997. PMID: 9224663 Review.
Cited by
-
A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation.Osteoporos Int. 2007 Apr;18(4):513-23. doi: 10.1007/s00198-006-0280-1. Epub 2006 Nov 30. Osteoporos Int. 2007. PMID: 17136486
-
[Absorptiometry].Radiologe. 2006 Oct;46(10):847-60. doi: 10.1007/s00117-006-1414-3. Radiologe. 2006. PMID: 17021911 Review. German.
-
Quality Control of DXA System and Precision Test of Radio-technologists.J Bone Metab. 2014 Feb;21(1):2-7. doi: 10.11005/jbm.2014.21.1.2. Epub 2014 Feb 28. J Bone Metab. 2014. PMID: 24707462 Free PMC article. Review.