Intratracheal pulmonary ventilation and continuous positive airway pressure in a sheep model of severe acute respiratory failure
- PMID: 9377918
- DOI: 10.1378/chest.112.4.1060
Intratracheal pulmonary ventilation and continuous positive airway pressure in a sheep model of severe acute respiratory failure
Abstract
Study objectives: Previously we have shown that optimal pulmonary gas exchange can be sustained at normal airway pressures in a model of severe acute respiratory failure (ARF), using intratracheal pulmonary ventilation (ITPV), with weaning to room air. In an identical model of ARF, we have now explored whether ITPV, combined with continuous positive airway pressure (CPAP), can sustain adequate ventilation, with weaning to room air.
Design: Randomized study in sheep.
Setting: Animal research laboratory at the National Institutes of Health.
Interventions: ARF was induced in 12 sheep, using mechanical ventilation at peak inspiratory pressure of 50 cm H2O, but excluding 5 to 8% of lungs. Sheep were then randomized into two groups: the CPAP-ITPV group (n=6), in which ITPV was combined with a novel CPAP system; and a control group (n=6) in which the same CPAP circuit was used, but without ITPV.
Measurements and results: All sheep in the CPAP-ITPV group were weaned to room air in 38.7+/-14 h. PaO2/fraction of inspired oxygen (FIO2) progressively increased from 108.8+/-43 to 355.7+/-93.1; PaCO2 remained within normal range; respiratory rate (RR) ranged from 18 to 120 breaths/min, and tidal volume (VT) was as low as 1.1 mL/kg. All sheep in the control group (CPAP alone) developed severe respiratory acidosis and hypoxemia after 4.8+/-4 h. PaO2/FIO2 decreased from 126.6+/-58.2 to 107.2+/-52.5 mm Hg, with a final PaCO2 of 166.8+/-73.3 mm Hg.
Conclusions: All sheep treated with CPAP-ITPV maintained good gas exchange without hypercapnia at high RR and at low VT, with weaning to room air. All control animals treated with CPAP alone developed severe hypercapnia, respiratory acidosis, and severe hypoxemia, and were killed.
Similar articles
-
Intratracheal pulmonary ventilation at low airway pressures in a ventilator-induced model of acute respiratory failure improves lung function and survival.Chest. 1998 Oct;114(4):1147-57. doi: 10.1378/chest.114.4.1147. Chest. 1998. PMID: 9792591
-
Comparison of intratracheal pulmonary ventilation and hybrid intratracheal pulmonary ventilation with conventional mechanical ventilation in a rabbit model of acute respiratory distress syndrome by saline lavage.Crit Care Med. 2000 Mar;28(3):774-81. doi: 10.1097/00003246-200003000-00028. Crit Care Med. 2000. PMID: 10752829
-
Intratracheal pulmonary ventilation in a rabbit lung injury model: continuous airway pressure monitoring and gas exchange efficacy.Crit Care Med. 2000 Jul;28(7):2480-5. doi: 10.1097/00003246-200007000-00049. Crit Care Med. 2000. PMID: 10921582
-
Review: artifical ventilation with positive end-expiratory pressure (PEEP). Historical background, terminology and patho-physiology.Eur J Intensive Care Med. 1976 Sep;2(2):77-85. doi: 10.1007/BF01886120. Eur J Intensive Care Med. 1976. PMID: 786640 Review.
-
The effect of respiratory frequency on pulmonary function during artificial ventilation. A review.Acta Anaesthesiol Scand. 1976;20(1):20-31. doi: 10.1111/j.1399-6576.1976.tb05005.x. Acta Anaesthesiol Scand. 1976. PMID: 773076 Review.
Cited by
-
Use of a Portable Mechanical Ventilator during Cardiopulmonary Resuscitation is Feasible, Improves Respiratory Parameters, and Prevents the Decrease of Dynamic Lung Compliance.Arq Bras Cardiol. 2023 Jul;120(7):e20220564. doi: 10.36660/abc.20220564. Arq Bras Cardiol. 2023. PMID: 37585896 Free PMC article. English, Portuguese.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical