The efficiency of adiabatic inversion for perfusion imaging by arterial spin labeling
- PMID: 9430351
- DOI: 10.1002/(sici)1099-1492(199706/08)10:4/5<216::aid-nbm468>3.0.co;2-u
The efficiency of adiabatic inversion for perfusion imaging by arterial spin labeling
Abstract
Velocity-driven adiabatic inversion is an attractive method for labeling arterial blood spins for quantitative perfusion imaging. To quantify perfusion and to optimize experimental parameters, an accurate estimate of labeling efficiency is required. We present theoretical and numerical methods to calculate the labeling efficiency over a wide range of experimental and physiologic parameters. The results are compared to experimental measurements in vivo. Inversion efficiency was found to be higher than previously assumed and relatively insensitive to flow velocity and the amplitude of the RF irradiation used for inversion. Assuming laminar flow, labeling efficiencies of greater than 90% are easily obtainable over a broad range of flow velocities. For applications where RF power deposition is a limiting factor such as at high field strengths, labeling efficiency can be maintained by reducing the labeling gradient. These results further illustrate the capability of adiabatic inversion to effectively label flowing blood.
Comment in
-
Current issues in functional MRI.NMR Biomed. 1997 Jun-Aug;10(4-5):157-9. doi: 10.1002/(sici)1099-1492(199706/08)10:4/5<157::aid-nbm492>3.0.co;2-4. NMR Biomed. 1997. PMID: 9430341 Review. No abstract available.
Similar articles
-
Inversion profiles of adiabatic inversion pulses for flowing spins: the effects on labeling efficiency and labeling accuracy in perfusion imaging with pulsed arterial spin-labeling.Magn Reson Imaging. 2002 Jul;20(6):487-94. doi: 10.1016/s0730-725x(02)00525-8. Magn Reson Imaging. 2002. PMID: 12361796
-
Quantitative analysis of adiabatic fast passage for steady laminar and turbulent flows.Magn Reson Med. 2002 Apr;47(4):709-19. doi: 10.1002/mrm.10122. Magn Reson Med. 2002. PMID: 11948732
-
Simple model of double adiabatic inversion (DAI) efficiency.Magn Reson Med. 2004 Oct;52(4):941-6. doi: 10.1002/mrm.20240. Magn Reson Med. 2004. PMID: 15389933
-
Tissue specific perfusion imaging using arterial spin labeling.NMR Biomed. 1994 Mar;7(1-2):75-82. doi: 10.1002/nbm.1940070112. NMR Biomed. 1994. PMID: 8068529 Review.
-
A technical perspective for understanding quantitative arterial spin-labeling MR imaging using Q2TIPS.Magn Reson Med Sci. 2015;14(1):1-12. doi: 10.2463/mrms.2013-0064. Epub 2014 Dec 15. Magn Reson Med Sci. 2015. PMID: 25500774 Review.
Cited by
-
Magnetization transfer effects on the efficiency of flow-driven adiabatic fast passage inversion of arterial blood.NMR Biomed. 2007 Dec;20(8):733-42. doi: 10.1002/nbm.1137. NMR Biomed. 2007. PMID: 17304639 Free PMC article.
-
An Inexpensive and Programmable RF Transmitter Setup for Two-coil CASL.Concepts Magn Reson Part B Magn Reson Eng. 2008 Oct 1;33B(4):228-235. doi: 10.1002/cmr.b.20127. Concepts Magn Reson Part B Magn Reson Eng. 2008. PMID: 19830266 Free PMC article.
-
Overview and Critical Appraisal of Arterial Spin Labelling Technique in Brain Perfusion Imaging.Contrast Media Mol Imaging. 2018 May 8;2018:5360375. doi: 10.1155/2018/5360375. eCollection 2018. Contrast Media Mol Imaging. 2018. PMID: 29853806 Free PMC article. Review.
-
Cerebral hemodynamics and pseudo-continuous arterial spin labeling considerations in adults with sickle cell anemia.NMR Biomed. 2017 Feb;30(2):10.1002/nbm.3681. doi: 10.1002/nbm.3681. Epub 2017 Jan 4. NMR Biomed. 2017. PMID: 28052565 Free PMC article.
-
Nonlinear magnetic field gradients can reduce SAR in flow-driven arterial spin labeling measurements.J Magn Reson. 2007 Mar;185(1):94-102. doi: 10.1016/j.jmr.2006.11.011. Epub 2006 Dec 18. J Magn Reson. 2007. PMID: 17175188 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical