[Three-dimensional thickness and volume measurements of the knee joint cartilage using MRI: validation in an anatomical specimen by CT arthrography]
- PMID: 9440900
- DOI: 10.1055/s-2007-1015574
[Three-dimensional thickness and volume measurements of the knee joint cartilage using MRI: validation in an anatomical specimen by CT arthrography]
Abstract
Purpose: In the present study we intended to validate knee joint cartilage volume and thickness measurements with MRI.
Methods: Ten fresh cadaver knees (age 29 to 64 yrs.) were sagittally imaged, using a fat-suppressed FLASH-3D sequence with a resolution of 2 x 0.31 x 0.31 mm3. Then, a contrast agent was injected and the specimens submitted to CT arthrography. From both modalities the patellar, femoral, and tibial cartilages were segmented semiautomatically and reconstructed three-dimensionally. The cartilage thickness was determined independently of the sectional plane, based on a "minimal distance algorithm".
Results: The volumes and the regional distribution patterns yielded a very high degree of similarity on direct comparison of both imaging modalities. The average volume error between MRI and CT was 3.8% (+/- 3.0%), the correlation 0.998, the slope of the regression line 1.04 and the gamma-intercept -80 mm3. The analysis yielded no significant differences between the two methods (Wilcoxon signed rank test, 5% level) in the patella, femur, medial, and lateral tibia.
Conclusion: The results suggest that, based on a fat-suppressed FLASH sequence with high resolution and three-dimensional concepts of digital image analysis, the cartilage volume and thickness can be analysed non-invasively and with high accuracy by MRI.
Similar articles
-
[Three-dimensional thickness and volume measurements of the knee joint cartilage by MR tomography: reproducibility in volunteers].Rofo. 1997 Dec;167(6):585-90. doi: 10.1055/s-2007-1015587. Rofo. 1997. PMID: 9465953 German.
-
Accuracy of cartilage volume and thickness measurements with magnetic resonance imaging.Clin Orthop Relat Res. 1998 Jul;(352):137-48. Clin Orthop Relat Res. 1998. PMID: 9678042
-
[Knee joint cartilage in magnetic resonance tomography. MR chondrovolumetry (MR-CVM) using fat-suppressed FLASH 3D sequence].Radiologe. 1995 Feb;35(2):87-93. Radiologe. 1995. PMID: 7740121 German.
-
[Imaging of articular cartilage].Acta Orthop Traumatol Turc. 2007;41 Suppl 2:32-42. Acta Orthop Traumatol Turc. 2007. PMID: 18180582 Review. Turkish.
-
[MRI and articular cartilage].J Radiol. 1998 May;79(5):391-402. J Radiol. 1998. PMID: 9757267 Review. French.
Cited by
-
Comparison of Micro-Computed Tomography and Clinical Computed Tomography Protocols for Visualization of Nasal Cartilage Before Surgical Planning for Rhinoplasty.JAMA Facial Plast Surg. 2019 May 1;21(3):237-243. doi: 10.1001/jamafacial.2018.1931. JAMA Facial Plast Surg. 2019. PMID: 30730533 Free PMC article.
-
Visualization of mouse spinal cord microscopic structures by use of ex vivo quantitative micro-CT images.Radiol Phys Technol. 2013 Jan;6(1):7-13. doi: 10.1007/s12194-012-0163-4. Epub 2012 Jun 24. Radiol Phys Technol. 2013. PMID: 22729620
-
3D micro-CT imaging of the postmortem brain.J Neurosci Methods. 2008 Jun 30;171(2):207-13. doi: 10.1016/j.jneumeth.2008.03.006. Epub 2008 Mar 27. J Neurosci Methods. 2008. PMID: 18462802 Free PMC article.
-
Quantitative measurement of cartilage morphology in osteoarthritis: current knowledge and future directions.Skeletal Radiol. 2023 Nov;52(11):2107-2122. doi: 10.1007/s00256-022-04228-w. Epub 2022 Nov 15. Skeletal Radiol. 2023. PMID: 36380243 Free PMC article. Review.
-
Accuracy and Reliability of Cone-Beam Computed Tomography for Linear and Volumetric Mandibular Condyle Measurements. A Human Cadaver Study.Sci Rep. 2017 Sep 20;7(1):11993. doi: 10.1038/s41598-017-12100-4. Sci Rep. 2017. PMID: 28931867 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical