[Failure of total uncemented trapeziometacarpal prosthesis. A multicenter study]
- PMID: 9453742
- DOI: 10.1016/s0753-9053(97)80004-x
[Failure of total uncemented trapeziometacarpal prosthesis. A multicenter study]
Abstract
The total trapeziometacarpal prosthesis that we have designed has been used clinically for five years. We thought it would be useful to revise failures of these prostheses in a multicentre study in order to identify the causes of the problems and to provide necessary modifications for improvement of the results. Eleven surgeons participated in this study, allowing the review of 24 failures out of a total of 188 implanted prosthesis (12.7%). One case of sepsis and three prosthesis dislocations were observed. In ten cases, we found that the cup was not correctly oriented, leading to a progressive loosening of the cup. In five cases, the shaft had sunk into the metacarpal which was associated with modification of the axis of the shaft in relation to the metacarpal in two cases. In three of these cases, the size of the shaft was too small. Twenty-three failures were reoperated. Metallosis of periprosthetic tissues was observed in 6 cases. This disadvantage has disappeared since treating the heads with ionic nitrogen implantation. A larger size of the same type of prosthesis was able to be implanted in 5 cases. A silicone prosthesis was used in 2 cases and a trapezectomy with or without ligamentoplasty was performed in 16 cases. The high failure rate related to a poor orientation of the cup has led us to increase the range of motion of the prosthesis in order to compensate for the possibly poor orientation of the cup during implantation.
Similar articles
-
[Radiological course of cemented and uncemented trapeziometacarpal prostheses].Ann Chir Main Memb Super. 1997;16(3):222-8. doi: 10.1016/s0753-9053(97)80005-1. Ann Chir Main Memb Super. 1997. PMID: 9453743 French.
-
Long-term results of the ABG-1 hydroxyapatite coated total hip arthroplasty: analysis of 111 cases with a minimum follow-up of 10 years.Orthop Traumatol Surg Res. 2009 Dec;95(8):579-87. doi: 10.1016/j.otsr.2009.10.001. Orthop Traumatol Surg Res. 2009. PMID: 19926546
-
Total modular wrist prosthesis: a new design.Scand J Plast Reconstr Surg Hand Surg. 2004;38(3):160-5. doi: 10.1080/02844310310023945. Scand J Plast Reconstr Surg Hand Surg. 2004. PMID: 15259675
-
[Wrist joint arthroplasty: results after 41 prostheses].Handchir Mikrochir Plast Chir. 2009 Jun;41(3):141-7. doi: 10.1055/s-2008-1039117. Epub 2009 Feb 2. Handchir Mikrochir Plast Chir. 2009. PMID: 19189248 German.
-
[The De La Caffinière trapezo-metacarpal prosthesis in rhizarthrosis of the thumb. Apropos of 20 cases surgically treated between 1978 and 1990].Ann Chir Main Memb Super. 1994;13(1):26-35. doi: 10.1016/s0753-9053(05)80354-0. Ann Chir Main Memb Super. 1994. PMID: 7511908 Review. French.
Cited by
-
Dual-mobility thumb carpometacarpal joint arthroplasty. Keypoints of surgical procedure for a satisfying outcome.J Hand Microsurg. 2024 Nov 30;17(1):100195. doi: 10.1016/j.jham.2024.100195. eCollection 2025 Jan. J Hand Microsurg. 2024. PMID: 39720004 Review.
-
Optimal Center of Rotation for Ball-And-Socket Thumb Carpometacarpal Arthroplasty Identified Using Three-Dimensional Kinematic Analysis: A Pilot Study.Front Bioeng Biotechnol. 2022 Jun 1;10:868150. doi: 10.3389/fbioe.2022.868150. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35721848 Free PMC article.
-
Pyrocardan Trapeziometacarpal Joint Arthroplasty-Medium-Term Outcomes.J Wrist Surg. 2020 Dec;9(6):509-517. doi: 10.1055/s-0040-1714685. Epub 2020 Aug 21. J Wrist Surg. 2020. PMID: 33282537 Free PMC article.
-
Treatment of the trapeziometacarpal osteoarthritis by arthroplasty with a pyrocarbon implant.Int Orthop. 2016 Jul;40(7):1465-71. doi: 10.1007/s00264-015-3016-z. Epub 2015 Oct 23. Int Orthop. 2016. PMID: 26493390