A comparison of methods for the measurement of hemodialysis access recirculation and access blood flow rate
- PMID: 9466503
A comparison of methods for the measurement of hemodialysis access recirculation and access blood flow rate
Abstract
The ability to accurately measure access recirculation (AR) is of importance because its presence indicates access dysfunction and may explain why a prescribed Kt/V (urea) has not been delivered. The ability to measure access flow (Qa) allows access monitoring and the detection of impending access dysfunction. AR can be measured by indicator dilution or conductivity tracer techniques. Qa calculation is simple if AR can be detected. The previous techniques are used while the patient's blood lines are reversed to induce AR, and the Krivitski equation gives: Qa = Qb [1-r/r] where Qb = dialyzer blood flow and r = proportion of AR induced. Three methods for AR and Qa measurements were directly compared: 1) ultrasound dilution (Transonics Hemodialysis Monitor, Transonics Systems Incorporated) (TRANS); 2) hematocrit dilution (Crit-Line Monitor, In-Line Diagnostics) (CRIT); and 3) differential conductivity (Hemodynamic Monitor, GAMBRO Healthcare Incorporated) (HDM). Patients were cannulated in a standard fashion and dialysis commenced with lines in normal configuration. A HDM test was performed and, if AR = zero, the lines were reversed to induce AR. HDM, TRANS, and CRIT tests for AR were next done in rapid succession for direct comparison. Each test was repeated three times in succession, the device in random order, to assess test repeatability. Qb was taken from the 1) dialysis machine pump, and 2) directly from TRANS and Qa calculated, using 1) and 2) AR results. In comparison to TRANS, AR results were virtually identical for HDM (TRANS AR = 1.04 HDM-AR + 0.02, r = 0.98, p = 0.0000), and good for CRIT (CRIT-AR = 0.84 TRANS-AR - 0.2, r = 0.81, p = 0.001), but CRIT underestimated the values. Repeatability was assessed by normalizing (%) the SD of repeated measurements; values were 7.5% (HDM), 9.1 % (TRANS), and 17.4% (CRIT). Qa value comparisons were similar (minimal r = 0.83) regardless of Qb source, but CRIT overestimated the value; repeatability data showed 10.6% (HDM), 13.0% (TRANS), and 25.2% (CRIT) (n ranged from 15-64). In summary, TRANS and HDM appear equal as far as accuracy and repeatability of measurements; CRIT results correlated well, but tended to underestimate AR and overestimate Qa, and was less reproducible.
Similar articles
-
Hemodialysis blood access flow rates can be estimated accurately from on-line dialysate urea measurements and the knowledge of effective dialyzer urea clearance.Clin J Am Soc Nephrol. 2006 Sep;1(5):960-4. doi: 10.2215/CJN.00810306. Epub 2006 Jul 6. Clin J Am Soc Nephrol. 2006. PMID: 17699313
-
CQI: hemodialysis vascular access flow monitoring.CANNT J. 1999 Summer;9(3):42-5. CANNT J. 1999. PMID: 15712471
-
Estimation of hemodialysis access blood flow rates by a urea method is a poor predictor of access outcome.ASAIO J. 1998 Nov-Dec;44(6):818-22. doi: 10.1097/00002480-199811000-00010. ASAIO J. 1998. PMID: 9831091
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
-
Theoretical and practical issues in recirculation; assessment of vascular access.EDTNA ERCA J. 1998 Apr-Jun;24(2):3-6. EDTNA ERCA J. 1998. PMID: 10392054 Review.
Cited by
-
Monitoring and Surveillance of Hemodialysis Access.Semin Intervent Radiol. 2016 Mar;33(1):25-30. doi: 10.1055/s-0036-1572548. Semin Intervent Radiol. 2016. PMID: 27013773 Free PMC article. Review.
-
Comparing non-invasive diagnostic methods for arteriovenous fistula stenosis: a prospective study.J Ultrasound. 2023 Sep;26(3):687-693. doi: 10.1007/s40477-022-00731-x. Epub 2022 Nov 1. J Ultrasound. 2023. PMID: 36319839 Free PMC article.
-
A meta-analysis of randomized clinical trials assessing hemodialysis access thrombosis based on access flow monitoring: where do we stand?Semin Dial. 2015 Mar-Apr;28(2):E23-9. doi: 10.1111/sdi.12342. Epub 2015 Jan 28. Semin Dial. 2015. PMID: 25644548 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials