Compensator transducer increases ease, accuracy, and rapidity of measuring changes in specimen birefringence with polarization microscopy
- PMID: 946822
- DOI: 10.1111/j.1365-2818.1976.tb02384.x
Compensator transducer increases ease, accuracy, and rapidity of measuring changes in specimen birefringence with polarization microscopy
Abstract
An instrument has been designed to improve substantially the efficiency and convenience of measuring specimen birefringence retardation (BR) with a conventional Brace-Köhler compensator. The design is based on the precise transduction of the angular position of the compensator's Vernier dial to an easily visible red-illuminated display on a digital voltmeter. The instrument display is accurate to within 0.1 degrees over a range of +/-25 degrees in the compensator vernier dial position which is similar to the precision obtainable with an ordinary vernier and the Brace-Köhler method of measuring BR of biological specimens. For convenience in data reduction the display can also be read directly in nanometre units, with a similar precision for specimen BR in the range of 5 nm or less which is typical of birefringent fine structures in living cells.
Similar articles
-
Quantitative studies on the polarization optical properties of living cells. I. Microphotometric birefringence detection system.J Cell Biol. 1981 Apr;89(1):115-20. doi: 10.1083/jcb.89.1.115. J Cell Biol. 1981. PMID: 7228896 Free PMC article.
-
New polarized light microscope with precision universal compensator.J Microsc. 1995 Nov;180(Pt 2):140-7. doi: 10.1111/j.1365-2818.1995.tb03669.x. J Microsc. 1995. PMID: 8537959
-
Simultaneous use of Interphako interference contrast and polarization microscopy in the study of microorganisms.Folia Microbiol (Praha). 2021 Apr;66(2):241-246. doi: 10.1007/s12223-020-00839-9. Epub 2020 Nov 25. Folia Microbiol (Praha). 2021. PMID: 33241462
-
Centrifuge polarizing microscope. I. Rationale, design and instrument performance.J Microsc. 2001 Mar;201(Pt 3):341-56. doi: 10.1046/j.1365-2818.2001.00850.x. J Microsc. 2001. PMID: 11240850 Review.
-
Enhanced polarizing microscopy as a new tool in aneuploidy research in oocytes.Mutat Res. 2008 Mar 12;651(1-2):131-40. doi: 10.1016/j.mrgentox.2007.10.013. Epub 2007 Nov 9. Mutat Res. 2008. PMID: 18160331 Review.
Cited by
-
Rapid rate of tubulin dissociation from microtubules in the mitotic spindle in vivo measured by blocking polymerization with colchicine.J Cell Biol. 1984 Sep;99(3):1066-75. doi: 10.1083/jcb.99.3.1066. J Cell Biol. 1984. PMID: 6470037 Free PMC article.
-
Control mechanisms of the cell cycle: role of the spatial arrangement of spindle components in the timing of mitotic events.J Cell Biol. 1983 Sep;97(3):877-86. doi: 10.1083/jcb.97.3.877. J Cell Biol. 1983. PMID: 6885924 Free PMC article.
-
Calcium-labile mitotic spindles isolated from sea urchin eggs (Lytechinus variegatus).J Cell Biol. 1980 Aug;86(2):355-65. doi: 10.1083/jcb.86.2.355. J Cell Biol. 1980. PMID: 7190569 Free PMC article.
-
Quantitative studies on the polarization optical properties of living cells. I. Microphotometric birefringence detection system.J Cell Biol. 1981 Apr;89(1):115-20. doi: 10.1083/jcb.89.1.115. J Cell Biol. 1981. PMID: 7228896 Free PMC article.
-
Functional implications of cold-stable microtubules in kinetochore fibers of insect spermatocytes during anaphase.J Cell Biol. 1980 Jun;85(3):853-65. doi: 10.1083/jcb.85.3.853. J Cell Biol. 1980. PMID: 7391142 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources