A new silent magnetic resonance imaging using a rotating DC gradient
- PMID: 9469717
- DOI: 10.1002/mrm.1910390221
A new silent magnetic resonance imaging using a rotating DC gradient
Abstract
A new approach to silent MR imaging using a rotating DC gradient has been explored and experimentally studied. Acoustic or sound noise has been one of the problems in examining patients, mainly due to the fast gradient pulsings in interaction with the main magnetic field. The sound noise has been noted to be proportionately louder as the magnetic field strength becomes larger. In this report, we describe a new imaging technique using a mechanically rotating DC gradient coil. The rotating DC gradient coil can effectively replace both phase encoding as well as readout gradient pulsings, and data obtained in this manner can provide a set of projection data that later can be used for projection reconstruction. With some interpolation techniques one can also perform conventional two-dimensional fast Fourier transform image reconstruction. The sound noise intensity compared with the conventional imaging technique, such as the spin-echo sequence, has been reduced down to about -20.7 dB or 117.5 times with this new technique. The experimental pulse sequence and its principle are described and images obtained by the new silent MR imaging technique are reported.
Similar articles
-
Determination of gradient magnetic field-induced acoustic noise associated with the use of echo planar and three-dimensional, fast spin echo techniques.J Magn Reson Imaging. 1998 Sep-Oct;8(5):1154-7. doi: 10.1002/jmri.1880080522. J Magn Reson Imaging. 1998. PMID: 9786155
-
Acoustic noise reduction in MRI using Silent Scan: an initial experience.Diagn Interv Radiol. 2014 Jul-Aug;20(4):360-3. doi: 10.5152/dir.2014.13458. Diagn Interv Radiol. 2014. PMID: 24808439 Free PMC article.
-
"Silent" MRI with soft gradient pulses.Magn Reson Med. 1999 Jul;42(1):6-10. doi: 10.1002/(sici)1522-2594(199907)42:1<6::aid-mrm2>3.0.co;2-d. Magn Reson Med. 1999. PMID: 10398943
-
[Fast and ultra-fast magnetic resonance tomography. Basic principles, pulse sequences and special properties].Radiologe. 1995 Dec;35(12):879-93. Radiologe. 1995. PMID: 8584631 Review. German.
-
Three-dimensional cardiac magnetic resonance imaging.Am J Card Imaging. 1993 Sep;7(3):195-208. Am J Card Imaging. 1993. PMID: 10146375 Review.
Cited by
-
Effect of Acoustic fMRI-Scanner Noise on the Human Resting State.Brain Topogr. 2023 Jan;36(1):32-41. doi: 10.1007/s10548-022-00933-w. Epub 2022 Dec 19. Brain Topogr. 2023. PMID: 36536080 Free PMC article.
-
Silent BOLD imaging.MAGMA. 2001 Oct;13(2):76-81. doi: 10.1007/BF02668155. MAGMA. 2001. PMID: 11502421
-
Silent echo-planar imaging for auditory FMRI.MAGMA. 2008 Sep;21(5):317-25. doi: 10.1007/s10334-008-0132-4. Epub 2008 Aug 21. MAGMA. 2008. PMID: 18716815
-
Software-based noise reduction in cranial magnetic resonance imaging: Influence on image quality.PLoS One. 2018 Nov 1;13(11):e0206196. doi: 10.1371/journal.pone.0206196. eCollection 2018. PLoS One. 2018. PMID: 30383774 Free PMC article.
-
Quiet PROPELLER MRI techniques match the quality of conventional PROPELLER brain imaging techniques.AJNR Am J Neuroradiol. 2015 Jun;36(6):1124-7. doi: 10.3174/ajnr.A4235. Epub 2015 Feb 12. AJNR Am J Neuroradiol. 2015. PMID: 25678482 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical