Rotating samples in FT-RAMAN spectrometers
- PMID: 9477576
- DOI: 10.1016/s1386-1425(97)00165-0
Rotating samples in FT-RAMAN spectrometers
Abstract
It is customary to rotate samples in Raman spectroscopy to avoid absorption or sample heating. In FT-Raman experiments the rotation is always shown (typically 30-60 rpm) because higher speeds are thought to generate noise in the spectra. In this article we show that more rapid rotation is possible. A tablet containing maleic acid and one made up of sub-millimetre silica particles with metoprolol succinate as active ingredient were rotated at different speeds, up to 6760 rpm. The FT-Raman spectra were recorded and studied. We conclude that it is perfectly acceptable to rotate samples up to 1500 rpm.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources