Neutral temperature range in incubators: performance of equipment in current use and new developments
- PMID: 9505136
- DOI: 10.1615/critrevbiomedeng.v25.i4-5.10
Neutral temperature range in incubators: performance of equipment in current use and new developments
Abstract
Low-birth-weight neonates should be nursed at thermoneutrality inside incubators. Thermoneutrality control is essential to enhance body growth and to reduce neonatal illnesses and mortality. Guidelines have been published to provide the thermoneutral range, but the recommendations did not always take into account all ambient and physiological parameters influencing thermoneutrality. In most marketed incubators, the heat supply is controlled through convective air flow (closed incubators) or through radiant power density (radiant warmer beds). The heating unit (on/off cycling or adjustable proportional control) is activated by an error signal calculated from the difference between a controlled temperature and a reference value preset by the clinician. The controlled variable can be either the incubator air or the skin temperature of the anterior abdominal region of the neonate. The neonate's size, thermal properties of the mattress and of incubator walls, air temperature and humidity, air velocity, incubator wall temperatures all influence the heat exchanges between the neonate and the surroundings, and, consequently, modify the obtention of thermoneutrality. Moreover, studies of the physiological mechanisms by which the neonate regulates body heat storage suggest that metabolic rate, behavior, vigilance level, nursing care, and heater control processes should also be taken into account. Little attention has been paid to these factors, and incubator performances are often disappointing. This article reviews the different factors that modify thermoneutral condition. An attempt is made to suggest new ways to design equipment incorporating these factors in algorithms controlling heater processes in order to reach the optimal thermal environment in which the neonate should be nursed.
Similar articles
-
Additional double-wall roof in single-wall, closed, convective incubators: Impact on body heat loss from premature infants and optimal adjustment of the incubator air temperature.Med Eng Phys. 2016 Sep;38(9):922-8. doi: 10.1016/j.medengphy.2016.05.010. Epub 2016 Jul 4. Med Eng Phys. 2016. PMID: 27387899
-
Skin derivative control of thermal environment in a closed incubator.Med Biol Eng Comput. 1997 Sep;35(5):521-7. doi: 10.1007/BF02525534. Med Biol Eng Comput. 1997. PMID: 9374058
-
The heat balance of small babies nursed in incubators and under radiant warmers.Early Hum Dev. 1982 Apr;6(2):131-43. doi: 10.1016/0378-3782(82)90100-1. Early Hum Dev. 1982. PMID: 7094851
-
Thermoregulation: incubators, radiant warmers, artificial skins, and body hoods.Clin Perinatol. 1991 Sep;18(3):403-22. Clin Perinatol. 1991. PMID: 1934849 Review.
-
Mathematical modelling of thermoregulation processes for premature infants in closed convectively heated incubators.Comput Biol Med. 2015 Feb;57:159-72. doi: 10.1016/j.compbiomed.2014.11.021. Epub 2014 Dec 9. Comput Biol Med. 2015. PMID: 25557201 Review.
Cited by
-
Skin Temperatures of Back or Neck Are Better Than Abdomen for Indication of Average Proximal Skin Temperature During Sleep of School-Aged Children.Front Psychiatry. 2020 Sep 18;11:494528. doi: 10.3389/fpsyt.2020.494528. eCollection 2020. Front Psychiatry. 2020. PMID: 33061911 Free PMC article.
-
Experimental and numerical studies on convective heat transfer in a neonatal incubator.Med Biol Eng Comput. 2002 Jan;40(1):114-21. doi: 10.1007/BF02347704. Med Biol Eng Comput. 2002. PMID: 11954698
-
Influence of thermal drive on central sleep apnea in the preterm neonate.Sleep. 2008 Apr;31(4):549-56. doi: 10.1093/sleep/31.4.549. Sleep. 2008. PMID: 18457243 Free PMC article.