. 1998 Apr;36(4):913-7.
doi: 10.1128/JCM.36.4.913-917.1998.

Novel recombinant-antigen enzyme immunoassay for serological diagnosis of syphilis

Affiliations

Novel recombinant-antigen enzyme immunoassay for serological diagnosis of syphilis

H Young et al. J Clin Microbiol. 1998 Apr.

Abstract

Enzyme immunoassay (EIA) is an ideal method for screening large numbers of patients for syphilis. We evaluated a novel immune-capture EIA (ICE Syphilis; Murex Diagnostics) that uses three recombinant Treponema pallidum antigens (TpN15, TpN17, and TpN47) and compared the results with those obtained by the native T. pallidum antigen EIA (Captia SelectSyph-G; Centocor) that we currently use for the serodiagnosis of syphilis. Specificity was evaluated by screening 1,184 unselected serum specimens in parallel by the ICE Syphilis and SelectSyph-G assays, while sensitivity was tested with a panel of 101 serum specimens containing antitreponemal antibodies (treated and untreated) from patients with various stages of infection. The specificity of the ICE Syphilis EIA (99.8%) on screening was significantly higher (P < 0.02) than that of the SelectSyph-G EIA (99.2%). The sensitivity of the ICE Syphilis EIA was significantly higher (P < 0.01) than that of the SelectSyph-G EIA on both initial (99 versus 91.4%) and repeat (100 versus 92.4%) testing. The ICE Syphilis EIA was also significantly more sensitive (P < 0.01) than the fluorescent treponemal antibody-abs (92.4%) but not the T. pallidum hemagglutination assay (97.1%). Sera containing antitreponemal antibodies gave a much higher antibody index (absorbance of test serum/kit cutoff) by the ICE Syphilis EIA than by the SelectSyph-G EIA. This combined with the overall high sensitivity makes the ICE Syphilis EIA an ideal test for excluding or detecting treponemal infection in human immunodeficiency virus (HIV)-infected patients. The ICE Syphilis EIA was positive with sera from all 15 HIV-infected patients in the study, whereas sera from 3 HIV-infected patients were negative by the SelectSyph-G EIA. We conclude that the high sensitivity and specificity of the ICE Syphilis EIA and its suitability for automation make it an ideal screening test.

PubMed Disclaimer

References

    1. Baker-Zander S A, Hook E W, Bonin P, Handsfield H H, Lukehart S A. Antigens of Treponema pallidum recognized by IgG and IgM antibodies during syphilis in humans. J Infect Dis. 1985;151:264–272. - PubMed
    1. Berkowitz K, Baxi L, Fox H E. False-negative syphilis screening: the prozone phenomenon, nonimmune hydrops, and diagnosis of syphilis during pregnancy. Am J Obstet Gynecol. 1990;163:975–977. - PubMed
    1. Chelala C. Syphilis in the USA: the good, and the bad news. Lancet. 1997;349:1675.
    1. Department of Health. Sexually transmitted diseases, England 1995. New cases seen at NHS genito-urinary medicine clinics. Statistical bulletin 1–5. London, United Kingdom: Department of Health; 1996.
    1. Gerber A, Krell S, Morenz J. Recombinant Treponema pallidum antigens in syphilis serology. Immunobiology. 1997;196:535–549. - PubMed

MeSH terms

LinkOut - more resources