Comparative Study
. 1998 Jan;44(1):92-5.

Evaluation of the Sysmex UF-100 automated urinalysis analyzer

Affiliations
  • PMID: 9550564
Comparative Study

Evaluation of the Sysmex UF-100 automated urinalysis analyzer

J Ben-Ezra et al. Clin Chem. 1998 Jan.

Abstract

Urinalysis is a high-volume procedure that currently requires significant labor to examine microscopic sediment. We evaluated the Sysmex UF-100 automated urinalysis analyzer for performing this task. Instrument accuracy was assessed by comparing continuous counts of microscopic elements from the UF-100 with ranges of cells (per low-power field or high-power field) from manual microscopy performed on centrifuged urines. Counts showed good agreement between methods (gamma statistic: 0.880-0.970) for all microscopic elements in 252 urine samples. Within-run imprecision of cell counts expressed as CV (mean cell count/microL) was for erythrocytes (RBC) 31% (5), 18% (50), 2.4% (800); for leukocytes (WBC) 14% (10), 11% (100), 8.5% (400); for squamous epithelial cells (SEC) 18% (5), 12% (30), 7.0% (100); for casts 45% (1), 17% (4); for bacteria 2-12% (entire range of 40-2500). Between-run imprecision on quality-control cell suspensions expressed as CV (mean cell count/microL) was for RBC 6.1% (50), 2.7% (256); for WBC 26.9% (54), 4.9% (228). Cells counted on dilution were 99.1% of expected for RBC, 102.0% for WBC, and 121.8% for bacteria. Carryover was <0.04% for RBC, <0.03% for WBC, <0.14% for SEC, <0.29% for bacteria. We conclude that the UF-100 can automatically perform reliable quantitative microscopic urinalysis in batches without operator interaction.

PubMed Disclaimer

Publication types

LinkOut - more resources