Adequacy of haemodialysis with cuffed central-vein catheters
- PMID: 9550658
- DOI: 10.1093/ndt/13.3.745
Adequacy of haemodialysis with cuffed central-vein catheters
Abstract
Cuffed central-venous haemodialysis catheters are emerging as an alternative permanent haemodialysis vascular access. There is limited data regarding the adequacy of dialysis with prolonged used of these catheters. We conducted a prospective study comparing three commonly used cuffed central-venous haemodialysis catheters: (1) PermCath, Quinton Instrument Co, Seattle; (2) Tesio, Med Comp, Inc, Harleysville PA; (3) VasCath Soft Cell, Bard Instrument Company, Toronto, and compared them with control patients dialysing with arteriovenous access (AV) access. We randomly assigned 64 patients who needed prolonged temporary vascular access to placement of one of three catheters. The control group comprised 222 patients dialysing simultaneously in the same units with AV access.
Methods: All patients were dialysed with identical machines and kidneys. Maximal effort were made with every catheter to optimize achievable blood flow. Catheters with mechanical problems were treated first with urokinase and then fibrin sheath catheter stripping. The mean blood flow was determined by averaging mean blood flows from 30 consecutive treatments. Reliability of catheter was defined as percentage of treatments that were performed at a median blood flow of 350 ml/min or above during these 30 treatments. Kt/V was measured monthly and calculated using the single-pool Daugirdas formula. Haemodialysis prescription were adjusted for Kt/V above 1.2. Recirculation was measured using two-needle low-blood-flow technique.
Results: The mean blood flows were (PermCath 383.6 ml/min, Tesio 396.3 ml/min, VasCath 320.4 ml/min). PermCath and Tesio had comparable mean blood flows and were significantly higher than VasCath (P<0.005). Reliability of catheters were (PermCath 86.9%, Tesio 81.6%, VasCath 42.3%). Tesio and PermCath were equally reliable and both were more reliable than VasCath (P<0.005). Had the target for reliability been 300 ml/min all three catheters would have been equally reliable. Negative arterial pressure in excess of 300 mmHg prevented faster blood flows in 98% of instances. None of the catheters performed as well as the control population with AV access (mean blood flow 437ml/min, reliability 96%, P<0.005). Recirculation rates were 3.7% for PermCath 3.9% for Tesio, and 4% for VasCath. All patients weighing less than 85 kg acheived a Kt/V of 1.2 with a 4-h treatment. For comparison purposes when Kt/V was normalized to a 70-kg patient the results were PermCath 1.42. Tesio 1.44, VasCath 1.19, AV access 1.64.
Summary: All three catheters are capable of providing adequate haemodialysis although large patients will need extended treatment times. The PermCath and Tesio provide blood flow and reliability superior to the VasCath. Blood flow is limited in all catheters by inflow, as evidenced by negative arterial pressure. All catheters had acceptable recirculation. AV access is superior in terms of blood flow and reliability to all tested catheters.
Similar articles
-
All currently used measurements of recirculation in blood access by chemical methods are flawed due to intradialytic disequilibrium or recirculation at low flow.Am J Kidney Dis. 1998 Dec;32(6):1046-58. doi: 10.1016/s0272-6386(98)70082-4. Am J Kidney Dis. 1998. PMID: 9856523
-
Successful use of central venous catheter as permanent hemodialysis access: 84-month follow-Up in lucania.Blood Purif. 2001;19(1):39-43. doi: 10.1159/000014477. Blood Purif. 2001. PMID: 11114576
-
Effective flow performances and dialysis doses delivered with permanent catheters: a 24-month comparative study of permanent catheters versus arterio-venous vascular accesses.Nephrol Dial Transplant. 2002 Jul;17(7):1286-92. doi: 10.1093/ndt/17.7.1286. Nephrol Dial Transplant. 2002. PMID: 12105254 Clinical Trial.
-
Recirculation, urea disequilibrium, and dialysis efficiency: peripheral arteriovenous versus central venovenous vascular access.Am J Kidney Dis. 1997 Apr;29(4):479-89. doi: 10.1016/s0272-6386(97)90328-0. Am J Kidney Dis. 1997. PMID: 9100035 Review.
-
Vascular access as a determinant of adequacy of dialysis.Semin Nephrol. 2005 Mar;25(2):96-101. doi: 10.1016/j.semnephrol.2004.09.016. Semin Nephrol. 2005. PMID: 15791561 Review.
Cited by
-
Successful Image-Guided Retrieval of an Embolized Fragment of a Fractured Haemodialysis Catheter Tip from the Pulmonary Artery.Case Rep Nephrol Dial. 2016 Feb 3;6(1):21-5. doi: 10.1159/000443728. eCollection 2016 Jan-Apr. Case Rep Nephrol Dial. 2016. PMID: 26998477 Free PMC article.
-
A review article: access recirculation among end stage renal disease patients undergoing maintenance hemodialysis.Nephrourol Mon. 2013 Spring;5(2):728-32. doi: 10.5812/numonthly.6689. Epub 2013 Mar 30. Nephrourol Mon. 2013. PMID: 23841034 Free PMC article.
-
Modifiable risk factors for early mortality on hemodialysis.Int J Nephrol. 2012;2012:435736. doi: 10.1155/2012/435736. Epub 2012 Jul 24. Int J Nephrol. 2012. PMID: 22888426 Free PMC article.
-
Interventions for treating central venous haemodialysis catheter malfunction.Cochrane Database Syst Rev. 2017 Oct 26;10(10):CD011953. doi: 10.1002/14651858.CD011953.pub2. Cochrane Database Syst Rev. 2017. PMID: 29106711 Free PMC article. Review.
-
Reporting of "dialysis adequacy" as an outcome in randomised trials conducted in adults on haemodialysis.PLoS One. 2019 Feb 5;14(2):e0207045. doi: 10.1371/journal.pone.0207045. eCollection 2019. PLoS One. 2019. PMID: 30721242 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous