Experience with fully 3D PET and implications for future high-resolution 3D tomographs
- PMID: 9572503
- DOI: 10.1088/0031-9155/43/4/007
Experience with fully 3D PET and implications for future high-resolution 3D tomographs
Abstract
The aim of this paper is to report on experience with 3D positron emission tomography (PET) in our institute where we have three 3D scanners, of which two operate exclusively in 3D mode (ECAT ART, EXACT 3D). Fully 3D PET requires attention to a number of factors which are not as problematic in 2D PET. Firstly, 3D tomographs designed for whole-body acquisition suffer from a large single-photon field of view, extending well beyond the coincidence field of view. Single photons from outside the coincidence field of view increase the dead time and random coincidence rates, and contribute scattered events. For brain studies, we have extended the lead side shielding at the ends of the tomograph to mitigate against these effects, and this has dramatically improved the count rate performance. This approach is not as effective for whole-body scanning. In addition, operating in 3D without septa necessitates new approaches to transmission scanning, as measurements using positron emitters such as 68Ge/68Ga have the unfavourable characteristics of high dead time and high scatter. Both of our fully 3D scanners use 137Cs for single-photon transmission measurements, although the data are treated differently. On the ECAT ART, a combination of physical and electronic collimation effectively reduces transmission scatter to acceptable levels. On the EXACT 3D physical collimation is not as readily implemented and therefore segmentation and reassignment of the histogrammed attenuation (mu) values is employed to produce unbiased attenuation correction factors in 3D. Many of the lessons learnt with these BGO (bismuth germanate) based tomographs will be applicable to the next generation of systems using faster detectors such as lutetium oxyorthosilicate (LSO).
Similar articles
-
The effect of activity outside the direct field of view in a 3D-only whole-body positron tomograph.Phys Med Biol. 1998 Apr;43(4):895-904. doi: 10.1088/0031-9155/43/4/017. Phys Med Biol. 1998. PMID: 9572513
-
Physical characteristics of the ECAT EXACT3D positron tomograph.Phys Med Biol. 2000 Sep;45(9):2601-18. doi: 10.1088/0031-9155/45/9/313. Phys Med Biol. 2000. PMID: 11008960
-
Single-photon transmission measurements in positron tomography using 137Cs.Phys Med Biol. 1995 Jul;40(7):1255-66. doi: 10.1088/0031-9155/40/7/008. Phys Med Biol. 1995. PMID: 7568381
-
PET imaging using gamma camera systems: a review.Nucl Med Commun. 1996 Sep;17(9):758-66. doi: 10.1097/00006231-199609000-00006. Nucl Med Commun. 1996. PMID: 8895903 Review.
-
New cardiac cameras: single-photon emission CT and PET.Semin Nucl Med. 2014 Jul;44(4):232-51. doi: 10.1053/j.semnuclmed.2014.04.003. Semin Nucl Med. 2014. PMID: 24948149 Review.
Cited by
-
Respiratory gating of cardiac PET data in list-mode acquisition.Eur J Nucl Med Mol Imaging. 2006 May;33(5):584-8. doi: 10.1007/s00259-005-0031-0. Epub 2006 Jan 17. Eur J Nucl Med Mol Imaging. 2006. PMID: 16416329
-
Acute induction of anxiety in humans by delta-9-tetrahydrocannabinol related to amygdalar cannabinoid-1 (CB1) receptors.Sci Rep. 2017 Nov 3;7(1):15025. doi: 10.1038/s41598-017-14203-4. Sci Rep. 2017. PMID: 29101333 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials