Gas leak and gastric insufflation during controlled ventilation: face mask versus laryngeal mask airway

Affiliations
Clinical Trial

Gas leak and gastric insufflation during controlled ventilation: face mask versus laryngeal mask airway

L M Ho-Tai et al. Can J Anaesth. 1998 Mar.

Abstract

Purpose: To compare two airway management techniques, face mask (FM) with oropharyngeal airway and laryngeal mask airway (LMA), with respect to the effectiveness of positive pressure ventilation and airway maintenance.

Methods: After induction of anaesthesia, two airway management techniques (FM or LMA) and three peak pressures (20, 25 and 30 cm H2O) were randomly applied during controlled ventilation in 60 patients. Data collected included inspiratory and expiratory volumes and presence of gastro-oesophageal insufflation. Leak was calculated by subtracting the expiratory from the inspiratory volume, expressed as a fraction of the inspiratory volume.

Results: Expiratory volumes (mean +/- SD) at 20, 25 and 30 cm H2O for LMA ventilation were 893 +/- 260, 986 +/- 276 and 1006 +/- 262 respectively, and for FM ventilation 964 +/- 264, 1100 +/- 268 and 1116 +/- 261. Leak fractions at 20, 25 and 30 cm H2O for LMA ventilation were 0.21 +/- 0.15, 0.24 +/- 0.18 and 0.26 +/- 0.18 respectively, and for FM ventilation 0.14 +/- 0.09, 0.14 +/- 0.09 and 0.12 +/- 0.08. The frequency of gastro-oesophageal insufflation was 1.6%, 5% and 5% for the LMA and 5%, 15% and 26.6% for the FM for ventilation pressures of 20, 25 and 30 cm H2O respectively which was greater with LMA use.

Conclusion: Ventilation was adequate in all patients using both techniques. Leak was pressure dependent and greater with LMA use. Most of the leak was vented to the atmosphere via the pharynx. Gastro-oesophageal insufflation was more frequent with ventilation using the face mask. LMA use with positive pressure ventilation would appear to be a better airway management method than the face mask.