A continuous-flow capillary mixing method to monitor reactions on the microsecond time scale

Affiliations

A continuous-flow capillary mixing method to monitor reactions on the microsecond time scale

M C Shastry et al. Biophys J. 1998 May.

Abstract

A continuous-flow capillary mixing apparatus, based on the original design of Regenfuss et al. (Regenfuss, P., R. M. Clegg, M. J. Fulwyler, F. J. Barrantes, and T. M. Jovin. 1985. Rev. Sci. Instrum. 56:283-290), has been developed with significant advances in mixer design, detection method and data analysis. To overcome the problems associated with the free-flowing jet used for observation in the original design (instability, optical artifacts due to scattering, poor definition of the geometry), the solution emerging from the capillary is injected directly into a flow-cell joined to the tip of the outer capillary via a ground-glass joint. The reaction kinetics are followed by measuring fluorescence versus distance downstream from the mixer, using an Hg(Xe) arc lamp for excitation and a digital camera with a UV-sensitized CCD detector for detection. Test reactions involving fluorescent dyes indicate that mixing is completed within 15 micros of its initiation and that the dead time of the measurement is 45 +/- 5 micros, which represents a >30-fold improvement in time resolution over conventional stopped-flow instruments. The high sensitivity and linearity of the CCD camera have been instrumental in obtaining artifact-free kinetic data over the time window from approximately 45 micros to a few milliseconds with signal-to-noise levels comparable to those of conventional methods. The scope of the method is discussed and illustrated with an example of a protein folding reaction.

PubMed Disclaimer

References

    1. Biochemistry. 1996 Apr 30;35(17):5538-49 - PubMed
    1. J Biol Chem. 1983 Mar 25;258(6):3655-60 - PubMed
    1. Proteins. 1996 Apr;24(4):413-26 - PubMed
    1. Biochemistry. 1996 Dec 24;35(51):16852-62 - PubMed
    1. Nat Struct Biol. 1997 Jan;4(1):44-50 - PubMed

Publication types

LinkOut - more resources